Seite 27 von 53 Seiten
LEVELMETER 2000
Important:
The MINILEVEL NT must be switched ON via <ON/OFF> button the
red LED must be blinking.
The values are on the MINILEVEL NT displayed in µm/m or in
Arcsec
The values can be displayed on the LEVELMETER 2000 in the
choice selected by the user.
Recommended display units are:
•
X.XXX mm/m
•
X°XX’XX’’
2.1.3.1
Start the LEVELMETER 2000
press, ON/MODE key. The measuring mode last used is now being restarted,
then a display appears.
2.1.3.2
Measurement absolute
(1 Instrument only )
press repeatedly, until the cursor is positioned under SENSOR
confirm selection. Display is flashing.
press repeatedly until the required channel, e.g. "A" is displayed
confirm. The address is being searched
press to confirm displayed address
Displayed mode changes to ABSOLUTE respectively the last used mode
2.1.3.3
Differential Measurement
absolute (2 Instruments) as 2.1.3.2, but
press until "A-B" appears