page 22 of 65 pages
3.4
S
ELECTION OF THE FILTER UNDER DIFFERENT MEASURING CONDITIONS
/
FILTER
A number of different predefined filters can be selected.
Description of the different filter types:
•
FILTER 1:
No filtering, no integration of the measuring values (T const. = 0 .33 sec.)
•
FILTER 2:
Floating average of 3 measuring values (T const. = 1 sec.)
•
FILTER 3:
Floating average of 15 measuring values (T variable = 0.33 … 5 sec.)
•
FILTER 4:
Floating average of 6 measuring values (T const. = 2 sec.)
•
FILTER 5:
Floating average of 15 measuring values (T const. = 5 sec.)
T: Response time when changing the position. For filter 3 the actual change of the measuring value
will define the number of values to be used for calculating the floating average. With a considerable
change the number of values will be reduced with minute fluctuations the number will be increased.
Filter type 3 is the factory setting when leaving WYLER AG.
With the key
select the menu item [FILTER] and
confirm this selection with
.
Using the keys
you can now select the filter type
desired and then confirm it with
.
The measuring instrument will return to the measuring mode.
3.5
A
BSOLUTE MEASUREMENT
/
R
ELATIVE MEASUREMENT
3.5.1
A
BSOLUTE MEASUREMENT
As a factory setting the BlueLEVEL -2D will be programmed for absolute measurement (default setting).
If this is not the case select the function [Absolute]. After confirming this function with the key
the instrument is ready for measurements in the mode "ABSOLUTE".
The measuring value corresponds to the
v alue of the BlueLEVEL-2D minus "ZERO - OFFSET"
Inclination X: 0.996mm/m
Inclination Y: 4.311mm/m
Summary of Contents for BlueLEVEL-2D
Page 2: ......
Page 37: ...page 37 of 65 pages APPENDICES A FLOAT CHARTS GENERAL A1 START THE INSTRUMENT ON OFF...
Page 38: ...page 38 of 65 pages A2 KEY REL ZERO...
Page 39: ...page 39 of 65 pages A3 KEY HOLD FREEZING OF A MEASURING VALUE HOLD...
Page 40: ...page 40 of 65 pages A4 MENU SELECTION...
Page 41: ...page 41 of 65 pages A5 ADJUSTMENT OF THE DISPLAY DISPLAY...
Page 42: ...page 42 of 65 pages A6 MEASUREMENT ABSOLUTE MODE ABSOLUTE...
Page 43: ...page 43 of 65 pages A7 MEASURING ABSOLUTE CONSIDERING THE ZERO OFFSET MODE ABSOLUTE ZERO...
Page 44: ...page 44 of 65 pages A8 MEASURING RELATIVE MODE REL ZERO...
Page 45: ...page 45 of 65 pages A9 ADJUSTMENT OF THE MEASURING UNIT UNIT...
Page 46: ...page 46 of 65 pages A10 SETTING OF LIMITS LIMITS...
Page 47: ...page 47 of 65 pages A11 ADJUSTMENT OF FILTER TYPES FILTER...
Page 48: ...page 48 of 65 pages A12 GROUPING OF INSTRUMENTS WITH WIRELESS TRANSMISSION JOIN...
Page 49: ...page 49 of 65 pages A13 LEAVING A GROUP OF INSTRUMETNS LEAVE...
Page 50: ...page 50 of 65 pages B FLOAT CHARTS OPTIONS B1 PIN CODE QUERY FOR ACCESS TO THE OPTIONS...
Page 51: ...page 51 of 65 pages B2 SET PIN CODE...
Page 52: ...page 52 of 65 pages B3 DISPLAY SETTINGS...
Page 53: ...page 53 of 65 pages B4 LOGARITHMIC SCALE ON OFF LOGSCALE ON OFF...
Page 54: ...page 54 of 65 pages B5 PROGRAMMABLE KEYS REL ZERO AND ZERO SELECT PROGRAMMABLE KEYS...
Page 55: ...page 55 of 65 pages B6 SWITCHING FUNCTIONS ON OR OFF FUNCTIONS ON OFF...
Page 57: ...page 57 of 65 pages B8 SWITCH THE WIRELESS TRANSMISSION ON OR OFF RADIO ON OFF...
Page 58: ...page 58 of 65 pages B9 SWITCH THE COMPENSATION OF GRAVITY ON OR OFF GRAVITATION COMPENSATION...
Page 59: ...page 59 of 65 pages B10 DISPLAY THE FIRMWARE VERSION VERSION FIRMWARE...
Page 60: ...page 60 of 65 pages B11 RESET TO FACTORY SETTINGS FACTORY RESET...
Page 61: ...page 61 of 65 pages B12 FUNCTION CHECK DEVICE CHECK...