Summary of Contents for 51870

Page 1: ... Verify removable battery pack is in place Plug charger jack into outlet EMERGENCY Off Push to stop all functions Twist clockwise to restore Manual EMERGENCY Lowering of Sling Bar Turn handle clockwise to lower sling bar Powered EMERGENCY Lowering of Sling Bar Push in with pointed object to lower sling bar Push and hold RAISE or LOWER button B to position the lift bar arm simultaneously Push and h...

Page 2: ... bar Close Aerial legs Secure Sling onto Patient see sling IFU Expand Legs As Needed A Position Lift Sling Bar B Transport Patient Raise Patient to Desired Position B Attach Sling to Sling Bar Secure sling loops onto hooks and ensure safety latches close C All Yellow Areas on Equipment Are Adjustable Handset Functions Push Pull and Turn UNLOCK WHEELS Close Legs A B B A ...

Page 3: ...800 255 3126 wyeastmedical com 51925 R06 Aerial Mobile Full Body Lift Operator s Manual Page 1 of 20 AERIAL Operator s Manual OPERATOR S MANUAL AERIAL MOBILE FULL BODY LIFT Model No 51870 ...

Page 4: ...es the right to confirm the existence of any claimed defect and to determine whether any damage was caused by neglect misuse unauthorized alteration or failure to follow maintenance instructions With respect to a claim covered by this warranty the defective part will be replaced or repaired at no charge provided the defective part is returned to Wy East The user has the responsibility for installa...

Page 5: ..._____________ 11 TECHNICAL SPECIFICATIONS _________________________________________________ 12 PERIODIC MAINTENANCE ____________________________________________________ 13 AFTER EVERY USE_______________________________________________________ 13 ONCE A MONTH_________________________________________________________ 13 ONCE A YEAR___________________________________________________________ 13 TWICE A...

Page 6: ...this manual with the Aerial Mobile Full Body Lift and refer to it as required Contents of this manual are subject to change without prior notice to users DEFINITIONS The words WARNING CAUTION and NOTE carry specific meanings and should be carefully reviewed WARNING The personal safety of the patient or user may be involved Disregarding this information could result in injury to the patient or user...

Page 7: ...d mobility Used in conjunction with proper sling or other assistive device the Aerial will transfer or reposition a patient with comfort and convenience The Aerial has a compact footprint allowing for greater maneuverability in tight spaces Intended for use in acute care long term care outpatient care and home care It is not recommended for use in a shower Prior to using the Aerial for transfer or...

Page 8: ...EPA registered Hard Surface Disinfectant on painted and plastic surfaces On all other surfaces use an EPA registered Hard Surface disinfectant wipe that does not contain sodium hypochlorite bleach 10 Warranty applies only if repairs or alterations are performed by an authorized technician PRIOR TO FIRST USE The battery must be charged for 24 hours in order to reach proper function and prolong the ...

Page 9: ...rm 2 Press to lower lift arm 3 Press to open legs 4 Press to close legs Legs Open Close Function 1 Press on handset to open legs 2 Press on handset to close legs Allows clearance for the caregiver to get around a chair wheelchair or commode Lift Arm Lift Lower Function 1 Press on handset to raise the patient 2 Press on handset to lower the patient Lift the patient to the desired height Raise sling...

Page 10: ...will ship in the lowest mast height position To adjust remove black bolt caps unbolt 9 16 hex head bolt and nut With the assistance of another person remove bolt and lift the mast and arm tube to the desired location Realign the bolt through the base frame and mast with 1 washer on either side of the base frame Tighten bolt and nut torqued to 30 foot pounds Replace black caps and test to make sure...

Page 11: ...ease see Fig 1 for connections Control Box Display Upgrade only Standard Display Cycles done number of fully loaded lifts on actuator Work done Amp power Overloads Number of overloads attempted Days between service Number of days since last service Icons shown when hand control button active Battery Information Battery State 100 50 Battery State 50 25 Battery State Battery needs charging less than...

Page 12: ...tton is disengaged in order to return unit to functionality To disengage turn the red button a half turn clockwise and spring loaded action will pop the button out Manual Emergency Down In the event there is not enough power to lower the patient rotate the red collar at the top of the actuator rod as shown The lift arm will slowly lower the patient Powered Emergency Down In the event the handset d...

Page 13: ... Fig 2 4 Place the sling loop over the sling bar hook and ensure the safety latch closes Repeat loop placement per the sling IFU Fig 3 5 Ensure that the rear WHEELS are UNLOCKED 6 Raise the patient to the desired position 7 Move the lift to the location of intended transfer 8 Follow the sling IFU when lowering the patient and removing the sling loops from the sling bar hooks 9 Once the sling is no...

Page 14: ...st be stored without exposure to direct sunlight PARAMETER SPEC SWL 600 lbs Device Part Number 51870 Unit of Measure 1 Each Battery Li Ion 25 9V 2 25Ah Number of Lifts per Charge 35 40 Charge Time 24 hrs fully discharged Total Weight 87 lbs Wheel Diameter Front 3 or 2 Wheel Diameter Rear 4 Turning Diameter 64 3 Leg Under Bed Stretcher Clearance 3 8 or 2 5 Standard Door Width Clearance 31 6 Lifting...

Page 15: ...ust ONCE A YEAR Perform load testing TWICE A YEAR Raise the lift arm to its max height and wipe the cylinder rod of the actuators clean EVERY TWO YEARS Replace the battery NOTE On the upgraded control box with a display reset the Needs Service indicator by pressing the raise and lower buttons on the handset simultaneously and hold for 5 seconds ELECTRONICS BATTERY CARE AND CAUTIONS Handle batterie...

Page 16: ...o exit the deep discharge mode before use If the battery is completely discharged the charging will be started at a very small rate to protect the battery This small charging rate is not sufficient to turn on the light in the charger and therefore the user may believe the system has not yet started Depending on the battery state it may take several hours to get to the normal charging state The ora...

Page 17: ...________________________ 5 C to 40 C Relative humidity______________________________20 to 90 30 C not condensing Atmospheric pressure______________________________________________70 to 106 kPa STORAGE Temperature_____________________________________________________ 10 C to 50 C Relative humidity______________________________20 to 90 30 C not condensing Atmospheric pressure_________________________...

Page 18: ...oom arm actuator will not lift Too much weight on the lift Check the weight of person being lifted Actuator rod has been damaged Check for visual signs of side impact Call for service Broken connection Check that all cables completely pushed into control box Boom arm actuator will not lower Check for obstructions Remove obstruction Actuator rod has been damaged Check for visual signs of side impac...

Page 19: ...rs of your Aerial Mobile Full Body Lift 2 Purchase date if available 3 Part number and description of the replacement part needed see Replacement Parts section on page 18 of manual 4 Shipping and billing instructions including complete address RETURNING ITEMS To return parts or goods contact Wy East Medical for authorization Contact information Telephone 503 657 3101 Email customerservice wyeastme...

Page 20: ...431 52432 51672 51671 51648 51661 51652 51919 51687 51687 51687 REPLACEMENT PARTS This manual provides maintenance and service information for the Aerial Mobile Full Body Lift If replacement parts are required contact WY EAST MEDICAL with the serial number of your Aerial Mobile Full Body Lift to ensure the accuracy of replacement parts ...

Page 21: ...800 255 3126 wyeastmedical com 51925 R06 Aerial Mobile Full Body Lift Operator s Manual Page 19 of 20 AERIAL Operator s Manual NOTES ...

Page 22: ... s Manual Page 20 of 20 AERIAL Operator s Manual CONTACT INFORMATION For questions or information regarding other Wy East Medical products contact Wy East Medical Corporation 16700 SE 120th Ave Clackamas OR 97015 USA 503 657 3101 800 255 3126 wyeastmedical com info wyeastmed com ...