WUNDER SA.BI.SRL EN USER MANUAL
RA(WU150) REV. 02- 10/01/2017
4
Guide and Statement of manufacturer-Electromagnetic Immunity
The electronic scales RA model is scheduled for operation in the electromagnetic environment specified
below.
the customer and the user should ensure that it is used in that environment.
Immunity test
Compliance
Guide to electromagnetic environment
Electrostatic
discharges (ESD)
IEC/EN61000 - 4 -2
6kV in contact
8kV in air
The floors should be made of wood, concrete or ceramic. If
the floors are covered in synthetic material, the relative
humidity should be maximum 30%.
Rapid
transients/burst
IEC/EN61000 - 4 - 4
2kV power supply
The power supply should be of the type used typically in
commercial or hospital environments.
Surge
IEC/EN61000 - 4 - 5
1kV
differential
mode
The power supply should be of the type used typically in
commercial or hospital environments.
Voltage dips, short
interruption and
voltage variation
IEC/EN61000 - 4 - 11
<5%UT for 0.5 cycle
40%UT for 05 cycle
70%UT for 25 cycle
<5%UT for 5 sec
The power supply should be of the type used typically in
commercial or hospital environments. Note=UT is the value
of the voltage of the feed.
Power frequency
Magnetic field
IEC/EN61000 - 4 - 8
3A/m
NOTE: UT is the value of the power supply voltage.
CAUTION
The medical device requires particular electromagnetic compatibility precautions and must be installed and
used according to the information provided in the accompanying documents.
Manufacturer's guide and declaration - Electromagnetic emissions
The electronic scales RA model can be used in the electromagnetic environment specified below. The
customer or the user of the seat must ensure that it is used in compliance with the specifications below.
Emissions test
Compliance
Guide to electromagnetic environment
Radiated/
conducted /
emissions
IEC/EN61000-4-6
3Vrms 150kHz to
80MHz(for
appliances that are
not life supporting)
Equipment in RF communications and mobile laptops should ot
be used near any part of the tooth, including cables, except
when they meet the recommended separation distance
calculated by the frequency applied to the transmitter.