Maintenance, cleaning, disposal, spare parts
24
Note
After being connected to the instrument for approx. 60
minutes the probe is ready for operation, then
it is essential
to recalibrate the probe
.
Recommendation:
For precision measurements, calibrate the probe the
following day once again.
11 Refill only in case of:
– Big air bubbles
– Air bubbles at the gold cathode
– Measuring with high water pressure
– Probe cannot be calibrated or is sensitive to
knocking.