TetraCon 700
Measuring / Operation
29
ba55320e05
11/2009
4.2
Application-dependent settings
The TetraCon 700 conductivity measuring cell is long-term stable. The
conductivity measuring cell is immediately ready for operation for its
authorized use in water/wastewater applications.
Adapting the cell
constant to the
installation location
In the case of special installation conditions, it may be necessary to
adapt the cell constant (due to the influence of the measuring
environment, e.g. of boundary fields).
In chapter 6 W
HAT
TO
DO
IF
... of this operating manual you will find the
correct values to set for some products of the WTW accessory program
that require a correction of the cell constant (if it is possible to give fixed
values at all). If necessary, special installation recommendations for the
TetraCon 700 can be found in the accessory operating manual.
Note
Details on how to determine and set the cell constant can be taken from
the operating manual of the measuring transmitter.