SensoLyt
®
ORP 900-P
General information
5
ba75993d03
06/2017
2.2
Commissioning
Scope of delivery
ORP electrodeSensoLyt
®
ORP 900-P
Operating manual
Commissioning
Prepare the electrode for measuring as follows:
Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
Connect the sensor to the meter.
The sensor is immediately ready to measure.
Types of connection:
– via a connecting cable (accessory) to a free IDS connector on the meter
or
– wireless via an IDS WLM-S adapter (accessory) to a WLM-capable
meter
or
– via an IDS sensor slot of a multi parameter probe
(connected to a meter)
Accessories for the connection of the SensoLyt
®
ORP 900-P-P sensor to
the meter are listed in chapter 7 W
EAR
PARTS
AND
ACCESSORIES
.
To open and close the IDS plug-in position please note the section 2.1
O
PENING
AND
CLOSING
THE
PLUG
CONNECTION
.
2.3
Calibrating, measuring
Conversion
to standard
hydrogen
electrode
U
H
= U
Meas
+ U
Ref
U
Ref
depends on the reference system and temperature and can be taken
from the following table (see also DIN 38404-6 for the system Ag/AgCl/
saturated KCl):
Please keep the watering cap. It is required for the electrode
to be stored. Always keep the watering cap clean.
Remove the watering cap for calibrating and measuring. More
detailed information is given in the operating manual of the meter.
with:
U
H
= ORP, referring to the normal hydrogen electrode
U
Meas
= Measured ORP
U
Ref
= Voltage of the reference system compared to the normal
hydrogen electrode