SensoLyt
®
ORP 900-P
Cleaning
7
ba75993d03
06/2017
5
Cleaning
Cleaning
procedures
6
Technical data
Contamination
Cleaning
Gross contamination on the junction
Carefully brush off contamination
under running water using a soft
toothbrush
Fats, oils, protein-containing sedi-
ments and similar substances
Remove with household washing-
up liquid
Shaft dimensions,
material
Shaft length
120 mm
Shaft diameter
12 mm
Shaft material
Glass
IDS plug
Synthetic materials: Glass fiber reinforced
Noryl, TPU, TPC-ET, POM, PVC, PEEK,
PBT
O-ring: FPM
Contacts gold-plated
General data
ORP electrode
Platinum ring
Reference electrolyte
Polymer electrolyte
Junction
2-hole junction
Shunt conduction element
Ag/AgCl
Temperature sensor
Integrated NTC 30 (30 k
at 25 °C / 77 °F)
Measurement and
application
characteristics
ORP measuring range
-1250.0 ... +1250.0 mV
Minimum immersion depth
25 mm
Allowed temperature range
0 ... 60 °C (32 ... 140 °F)
Precision of the temperature
sensor
± 0.2 K
Response time t
99
of the
temperature sensor
< 130 s
Allowed pH range of the
measuring medium
4 ... 12
Typical application
Field