29
8. Analytical Quality Assurance (AQA)
– In the PhotoCheck standards submenu, call up the
output menu item.
Select the output medium:
l
to display
l
to printer/PC (serial interface).
– Select with
3
– Confirm with
1
to start the download.
At least 1 standard must still be stored to be able to
perform the AQA check function (meter monitoring).
– In the PhotoCheck standards submenu, call up the
erase menu item.
Displays the stored PhotoCheck standards:
– Select with
3
– Quit via return
– Erase with
1
.
8.2.2 Download of PhotoCheck standards
AQA check meter
AQA1
26.08.97
13:19
AQA interval
12 weeks
test sol.
unit
theor. val.
tolerance
AQA date
445-1
A
0.200
0.020
26.08.97
Example: Report output
8.2.3 Erasing PhotoCheck standards