What to do if...
pHotoFlex
®
STD
64
ba75958e07
11/2020
6
What to do if...
6.1
General errors
Display,
LoBat
Meter does not
react to keystroke
RS232 interface
does not react
Error message,
Error
0, 8, 16, 16384
Cause
Remedy
– The batteries or battery pack
are largely depleted
– Insert new batteries
– Charge the battery pack (see
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<START/ENTER>
and
<PRT>
keys simultaneously.
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<START/ENTER>
and
<PRT>
keys simultaneously.
Cause
Remedy
– Instrument error
– Repeat measurement
– Meter defective,
send meter to WTW for repair and
quote the error number