Operation
pH 3310
46
ba75802e02
12/2008
Continuing with two-
point calibration
10
Thoroughly rinse the combination electrode with distilled water.
11
Immerse the electrode in buffer solution 2.
12
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
▲
><
▼
>
.
13
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
14
Wait for the measurement with stability control to be
completed.
Set the nominal buffer value for the measured temperature with
<
▲
><
▼
>
.
15
Take over the calibration value for the slope with
<ENTER>
.
The calibration is completed as a two-point calibration.
The calibration record is displayed.
pH
10.000
24.8 °C
ConCal
2
22.09.2008
08:00
Slope