
MIQ/2-MOD
Interface for the Modbus RTU
15
ba75973e01
07/2012
3.4
Setting the Modbus interface parameters
The setting of the Modbus interface parameters (RS 485 interface) is
done in the menu,
MODBUS Konfiguration
.
Settings
Proceeding:
setting the interface
parameters
Please observe the operating manual of the connector.
Setting
Selection/Values
Description
Device Address
1 ... 247
Device address of
the IQ S
ENSOR
N
ET
Baud rate
1200
2400
4800
9600
19200
38400
57600
Baud rate
Parity
None
Even
Odd
Parity
None
(2 stop bits)
Even
(1 stop bit)
Odd
(1 stop bit)
1
Using <S>, open the
Settings
menu.
2
Press <
> and <OK> to select and confirm the menu
item,
Settings bus interfaces
.
3
Press <
> and <OK> to select and confirm the bus
interface.