3.6
Installation of the MIQ modules at the installation location
3.6.1 General information
The IQ S
ENSOR
N
ET
system has a comprehensive program of mount
-
ing accessories, which can be used to adapt the installation to the most
varied requirements.
Note
MIQ modules installed outside must always be protected by a sun
shield against the effects of the weather (snow, ice and direct solar ra
-
diation). Otherwise, malfunctions can result. Always mount
MIQ modules vertically. Do not under any circumstances install
MIQ modules without rain protection with the lid facing upwards (dan
-
ger of retained humidity and penetration of moisture).
Installation options
The most important types of installation for MIQ modules and module
stacks are described in the following chapters:
Mounting on a mounting stand with the SSH/IQ sun shield:
The SSH/IQ sun shield provides enough space for one unit of up to
three stacked MIQ modules and a docked terminal (section 3.6.2).
Mounting on the SD/K 170 sun shield
The SD/K 170 sun shield provides enough space for a single
MIQ module with a docked terminal. The sun shield can be mounted
on round or square section pipes (e.g. rails) with the aid of the MR/
SD 170 mounting kit (section 3.6.3).
Wall mounting:
The back MIQ module of a module stack or a single MIQ module is
permanently screwed to a wall. For wall mounting, use the WMS/IQ
mounting set (see chapter 12 A
Panel mounting:
The front MIQ module of a module stack or a single MIQ module is
installed in the aperture of a panel. The aperture dimensions are
138 x 138 mm + 1 mm (maximum thickness 15 mm) according to
DIN 43700 or IEC 473 (section 3.6.4).
Top hat rail mounting:
The back MIQ module of a module stack or a single MIQ module is
mounted on a 35 mm top hat rail with the aid of a bracket in accor
-
dance with
EN 50022, e.g. in a control cabinet. The connection can be released
again with one simple movement (section 3.6.5).
Caution
No contact base may be mounted on the back of the module (danger
of short-circuit!) if the module is mounted on a wall, a sun shield, or a
top hat rail.
Summary of Contents for IQ SENSOR NET System 2020 XT
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 8: ...Contents System 2020 XT CONNECT 0 6 ba75970e01 10 2012...
Page 24: ...Overview System 2020 XT CONNECT 1 16 ba75970e01 10 2012...
Page 28: ...Safety System 2020 XT CONNECT 2 4 ba75970e01 10 2012...
Page 123: ...System 2020 XT CONNECT Settings setup 5 11 ba75970e01 10 2012 Fig 5 10 Edit list of sensors...
Page 138: ...Settings setup System 2020 XT CONNECT 5 26 ba75970e01 10 2012...
Page 152: ...Network connection System 2020 XT CONNECT 6 14 ba75970e01 10 2012...
Page 178: ...PROFIBUS connection option System 2020 XT CONNECT 8 12 ba75970e01 10 2012...
Page 180: ...Maintenance and cleaning System 2020 XT CONNECT 9 2 ba75970e01 10 2012...
Page 194: ...What to do if System 2020 XT CONNECT 10 14 ba75970e01 10 2012...
Page 204: ...Accessories and options System 2020 XT CONNECT 12 2 ba75970e01 10 2012...
Page 212: ...Lists IQ SENSOR NET System 2020 XT CONNECT 13 8 ba75970e01 10 2012...
Page 215: ......