What to do if...
pH 720
34
Buffer solutions:
– Incorrect buffer solutions
– Change calibration
procedure
– Buffer solutions too old
– Only use once.
Note the shelf life
– Buffer solutions depleted
– Change solutions
Cause
Remedy
pH electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
Sample:
– pH value not stable
– Measure with air
excluded if necessary
– Temperature not stable
– Adjust temperature if
necessary
Ele sample:
– Conductivity too low
– Use suitable electrode
– Temperature too high
– Use suitable electrode
– Organic liquids
– Use suitable electrode
Cause
Remedy
– Batteries almost depleted
– Replace batteries
(see section
5.1 M
AINTENANCE
)
No stable
measured value
LoBat