
Operation
pH 720
22
Note
If the slope (
SLO
) is indicated on the display, you can
change the unit of the slope with
28
.
The display in % refers to the Nernst slope, 59.2 mV/pH
(100 x the slope determined/Nernst slope).
5 Thoroughly rinse the electrode with distilled water.
6 If necessary, set the temperature of the second buffer
solution by pressing
28
.
7 Submerse the electrode in the second buffer solution.
8 Press the
g
key.
AR flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized, AR disap-
pears.
The sensor symbol shows the electrode evaluation
after the two-point calibration.
The value of the slope (mV/pH) appears on the dis-
play.
9 Press the
g
key.
The value of the asymmetry (mV) appears on the dis-
play.
10 To return to the measuring mode: Press the
m
key
or continue with three-point calibration.
Summary of Contents for inoLab pH 720
Page 1: ......
Page 4: ...Contents pH 720 4...
Page 8: ...Overview pH 720 8...
Page 12: ...Safety pH 720 12...
Page 36: ...What to do if pH 720 36...
Page 40: ...Technical Data pH 720 40...
Page 46: ...46 Lists pH 720...
Page 47: ......