What to do if...
147
6
What to do if...
6.1
pH system messages
Cause
Remedy
pH electrode:
– Not connected
– Connect electrode
– Air bubbles in front of the
diaphragm
– Remove air bubbles
– Air in the diaphragm
– Extract air or
moisten diaphragm
– Cable broken
– Replace electrode
– Gel electrolyte dried out
– Replace electrode
Cause
Remedy
pH electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
– Moisture in the plug
– Dry plug
– Electrolyte obsolete
– Replenish electrolyte or
replace electrode
– Electrode obsolete
– Replace electrode
– Electrode broken
– Replace electrode
Measuring instrument:
– Incorrect calibration
procedure
– Select correct
procedure
– Incorrect solution tempera-
ture (without temp. probe)
– Set up correct
temperature
– Socket damp
– Dry socket
Error message,
OFL
OFL
OFL
OFL
Error message,
E3
E3
E3
E3