Operation
104
Calibration protocol
The calibration protocol contains the calibration data of the
current calibration. You can call up the calibration protocol
by outputting the data storage (section 4.7.3).
Note
You can automatically print out a calibration protocol after
the calibration. To do so, switch on the printer before the
calibration (Print LED lights up green). After a valid
calibration, the protocol is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.99
14:19
Device No.: 12345678
CALIBRATION pH
Cal Time: 01.03.99 / 15:20
Cal Interval:
7d
AutoCal DIN
Tauto
Buffer 1
1.679
Buffer 2
4.008 *
Buffer 3
6.865
Buffer 4
9.180 *
C1
174.1mV
25.0°C
C2
-133.3mV
25.0°C
S1
-59.4 mV/pH
ASY1
-
4 mV
Probe:
+++