Operation
132
4.8
Data transmission
You can use the following options to transmit data:
l
One of the following options:
– The AutoStore function (see page 127) is used to
periodically (Int 1 storage interval) save measured
values internally and output them on the interface.
– The data transmission interval
(Int 2) function is used to
periodically output measured values to the interface
(see below).
l
The Output data store function (see page 129) is used to
output calibration data and stored measured values to the
interface.
l
The analog recorder output (see page 134) is used to out-
put measured values as voltages.
l
Data can be transmitted bidirectionally (see page 136)
using the KOM pilot communication package (accessory).
4.8.1 Data transmission interval (Int 2)
The interval for the data transmission (Int 2) determines the
time interval between automatic data transmissions. After
the time interval expires, the current data record is transmit-
ted to the interface.
Note
The setting of the data transmission interval (Int 2) only has
an effect when the storage interval (AutoStore Int 1 function)
is switched off.
Setting the
data transmission
interval
The interval is set to OFF in the factory.
To start the data transmission, set up an interval (5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min):
1 Press and hold down the
g
key.