Determining the cell
constant
For this calibration procedure, the
Measuring cell
setting must be set to
cal in the
Measurement
menu. Proceed as follows to determine the cell
constant:
Display
Calibration
record
Cell constant [cm
-1
]
+++
within the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 0.880 cm
-1
Error
Error
outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 0.880 cm
-1
Eliminate the error according to
chapter 6 W
HAT
TO
DO
IF
...
1
Connect a conductivity measuring cell to the measuring instru-
ment.
2
In the measured value display, select the conductivity parame-
ter with
<M>
.
3
Start the calibration with
<CAL>
.
The cell constant that was calibrated last is displayed.
4
Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
5
Start the measurement with
<ENTER>
.
The [AR] status indicator is displayed. The measured parame-
ter flashes.
6
Wait for the end of the AutoRead measurement or accept the
calibration value with
<ENTER>
.
The calibration record is displayed.
ϰ
0.479 1
/cm
25.1
°C
15.03.2014
10:00