SI-BF-100
World Precision Instruments
9
Fig. 14—Line up the screw holes as illustrated here.
8. Insert the screws and tighten them down to re-mount the fi lter holder on the
face of the Biofl uorometer chassis.
Fig. 15—The fi lter holder has been reinstalled.
Adding an LED Module
Each
SI-BF-100
is confi gured for its intended application. For example, the
intracellular calcium (Fura-2) measurement system uses three LED modules and
one photomultiplier. The
Biofl uorometer
can be confi gured with up to seven LED
modules and two photomultipliers. You can add additional LED modules, if required.
1. Turn off the power to the
SI-BF-100
and unplug the instrument. Then, use a
Phillips screw driver to remove the four screws on the top of the Biofl uorometer
box.
2. Lift the lid and remove it to reveal the circuitry inside. The seven LED ports are
located in the front of the box. The are labeled 0–6 as shown in Fig. 16.