14
World Precision Instruments
3. Press
Date Format
. Choose your preferred format and press
Return
.
Fig. 18—(Left) Set the date and press Return
Fig. 19—(Right) Choose the date format and press Return
4. Press
Set Time
. Set the appropriate time and press
Return
.
5. Press
Time Format
. Choose 12 hour or 24 hour and press
Return
.
Fig. 20—(Left) Set the time and press Return
Fig. 21—(Right) Choose the time format and press Return
Probe Null
1. Press
Probe Null
on the
Setup
page to access the
NULL Probe Process
screen.
Fig. 22—(Left) Null probe process.
Fig. 23—(Right) The potential difference offset shows on the main screen under the reading.
2. Place the electrodes in a saline media.
3. Push the
NULL
button. Wait for the second message stating ”Solution value 0.0”
before returning. This action removes any voltage offsets on the electrodes and
neutralize the offset charge on V
1
-V
2.
4. Select the
Enable/Disable
radio button to use the offset.
5. Press
Return
. The potential difference offset shows on the main screen under the
reading. An electrode check should show 0.0 mV. The saved data file reflects the
offset.