WPI Biological Signals Amplifier Instruction Manual Download Page 9

Biological Signals Amplifier

World Precision Instruments 

5

 NOTE

: If the recorded output signals for the 10mmHg and 100mmHg settings 

are not 100mVDC and 1VDC respectively, recalibrate the pressure sensor to the 
HG1 module using the directions provided below.

6.  Once the calibration of the sensor and amplifier have been verified, prepare to 

attach the pressure sensor to the cannula or catheter. Purge all lines to remove 
any bubbles in the perfusion fluid lines. When ready, adjust the positions of 
the stopcocks on the sensor lines, and start recording changes in pressure by 
turning the 4-way program switch on the front of the HG1 module to 

MEAS

.

7. If a second pressure recording is required, connect a second pressure sensor to 

IN2

 input socket of the HG2 amplifier module. Connect the 

OUT2

 output socket 

of the HG2 module to another analog input on the data recorder using a similar 
output cable.

8. Check the baselines and calibration of the HG2 amplifier and it sensor using the 

same procedures used on HG1.

Calibrating the Pressure Sensor to HG Module

1.  Attach the pressure sensor and the recording system to the HG pressure module 

as described in the previous section.

2.  Turn the 4-way program switch on the front panel of HG amplifier to 

GND

. Set 

the 

BALANCE

 potentiometer to the middle position. Turn on the amplifier. A 

signal of zero (0) volts is sent by the amplifier to the data recorder. If necessary, 
set the baseline of the recording channel to zero (0) using the controls of the 
data recording system.

3.   Attach a certified manometer or blood pressure meter to the input port of the 

pressure sensor. Set the pressure in the manometer or blood pressure meter to 
zero(0) mm Hg

4.   Turn the 4-way program switch to 

MEAS

. If the baseline level on the recording 

channel changes when this switch is moved from 

GND

 to 

MEAS

, adjust the 

LOW

 trimmer-potentiometer on the front panel of the module until the baseline 

on the recording channel is set to zero (00). Remember that the manometer/BP 
meter has to be set at 0mmHg throughout this step. Return the 4-way program 
switch to 

GND

 before the next step.

5.  Set the manometer/BP meter to 

generate a pressure of 100mmHg. 
Turn the 4-way program switch 
to 

MEAS

. The baseline level on 

the recording channel should 
read 1VDC for the 100mmHg of 
pressure. If the value is different, 
adjust the 

HIGH

 potentiometer 

until the voltage level on the 
recording channel reads 1VDC.

Fig. 5—(Right) The HG pressure sensor.

Summary of Contents for Biological Signals Amplifier

Page 1: ...Biological Signals Amplifier for isolated heart Langendorff Neely and perfused organ systems World Precision Instruments INSTRUCTION MANUAL Serial No _____________________ 070212 www wpiinc com...

Page 2: ......

Page 3: ...prior written permission of World Precision Instruments Inc CONTENTS ABOUT THIS MANUAL 1 INTRODUCTION 1 Parts List 2 Unpacking 2 INSTRUMENT DESCRIPTION 3 Setting Up HG Pressure Amplifiers and Sensors...

Page 4: ...iv World Precision Instruments...

Page 5: ...y perfusate pressures temperature and ECG signals or surface potentials from isolated perfused heart Langendorff and working heart Neely preparations The system can also be used to measure perfusion p...

Page 6: ...at sales wpiinc com or 941 371 1003 to speak with one of our representatives regarding the type of electrodes needed to record the biopotentials of interest Unpacking Upon receipt of this instrument m...

Page 7: ...ls Amplifier World Precision Instruments 3 INSTRUMENT DESCRIPTION Fig 2 The front panel has the operation switches and indicators Fig 3 The back panel has the input and output sockets along with the m...

Page 8: ...cket of the HG1 pressure amplifier Connect the other end of the output cable to one of the analog inputs of the data acquisition unit 3 Before the transducer is connected to a perfusion cannula or cat...

Page 9: ...pressure module as described in the previous section 2 Turn the 4 way program switch on the front panel of HG amplifier to GND Set the BALANCE potentiometer to the middle position Turn on the amplifi...

Page 10: ...log inputs of the data acquisition unit 3 Turn the 3 way program switch to GND Place the temperature probe on or into the area where the temperature needs to be measured Turn on the amplifier unit The...

Page 11: ...ler fluid Place the tip of the probe as close to the bulb of the thermometer as possible without touching it 4 Turn the 3 way program switch to the MEAS position Wait until the temperature reading on...

Page 12: ...of interest The type of electrodes plate pellet needle wire etc determine the type of lead wires needed to connect the electrodes to the input cable of the ECG amplifier and the preparation required...

Page 13: ...ding channel Turn the 3 way program switch to 1mV The signal on the recording channel should show a deflection of 1mVDC which is the nominal voltage used to calibrate ECG amplifiers If needed use the...

Page 14: ...10 World Precision Instruments...

Page 15: ...ct indirect special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused d...

Page 16: ...44 0 1462 424700 Fax 44 0 1462 424701 E mail wpiuk wpi europe com Germany Zossener Str 55 10961 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading...

Reviews: