WPI Biological Signals Amplifier Instruction Manual Download Page 12

World Precision Instruments

Setting up the ECG Amplifier

The ECG amplifier is designed to record the ECG signals from as many as three 
different positions (leads I, II, III) on an animal using standard electrodes or surface 
potentials from the muscle fibers of an isolated heart using wire electrodes. This 
amplifier is appropriate for use on species in size from rats to dogs. The amplifier 
also has an internal rate meter which allows the heart rate to be recorded as ECG 
signals or surface potentials are also recorded.

If you are interested in recording ECG signals or other types of biopotentials, contact 
WPI at [email protected] or 941.371.1003 to speak with one of our representatives 
regarding the type of electrodes that would be best for recording the biopotentials of 
interest. The type of electrodes (plate, pellet, needle, wire, etc.) determine the type of 
lead wires needed to connect the electrodes to the  input cable of the ECG amplifier, 
and the preparation required to attach the electrodes to the animal and promote 
conduction of the signal.

 

Fig. 8—ECG Amplifier, front and back.

If the ECG amplifier is going to be used:
1.  Plug the connector on the end of the ECG input cable into the 

ECG-IN

 15-pin 

Canon (D-sub) socket on the back panel of the amplifier.

2.  Plug the connector on the end of the ECG output cable to the 

ECG-OUT

 9-pin 

Canon (D-sub) socket below the 

ECG-IN

 socket. Connect the other ends of the 

output cable, which are labeled with numbers that correspond to standard ECG 
limb leads [I (1), II (2), III (3)] to the analog inputs on the data recorder that are 
needed.

 

TIP

: If only one pair of electrodes are needed to record a surface potential from 

heart muscle, then only one of the leads (I, II, or III) needs to be connected to 
the data recorder.

Summary of Contents for Biological Signals Amplifier

Page 1: ...Biological Signals Amplifier for isolated heart Langendorff Neely and perfused organ systems World Precision Instruments INSTRUCTION MANUAL Serial No _____________________ 070212 www wpiinc com...

Page 2: ......

Page 3: ...prior written permission of World Precision Instruments Inc CONTENTS ABOUT THIS MANUAL 1 INTRODUCTION 1 Parts List 2 Unpacking 2 INSTRUMENT DESCRIPTION 3 Setting Up HG Pressure Amplifiers and Sensors...

Page 4: ...iv World Precision Instruments...

Page 5: ...y perfusate pressures temperature and ECG signals or surface potentials from isolated perfused heart Langendorff and working heart Neely preparations The system can also be used to measure perfusion p...

Page 6: ...at sales wpiinc com or 941 371 1003 to speak with one of our representatives regarding the type of electrodes needed to record the biopotentials of interest Unpacking Upon receipt of this instrument m...

Page 7: ...ls Amplifier World Precision Instruments 3 INSTRUMENT DESCRIPTION Fig 2 The front panel has the operation switches and indicators Fig 3 The back panel has the input and output sockets along with the m...

Page 8: ...cket of the HG1 pressure amplifier Connect the other end of the output cable to one of the analog inputs of the data acquisition unit 3 Before the transducer is connected to a perfusion cannula or cat...

Page 9: ...pressure module as described in the previous section 2 Turn the 4 way program switch on the front panel of HG amplifier to GND Set the BALANCE potentiometer to the middle position Turn on the amplifi...

Page 10: ...log inputs of the data acquisition unit 3 Turn the 3 way program switch to GND Place the temperature probe on or into the area where the temperature needs to be measured Turn on the amplifier unit The...

Page 11: ...ler fluid Place the tip of the probe as close to the bulb of the thermometer as possible without touching it 4 Turn the 3 way program switch to the MEAS position Wait until the temperature reading on...

Page 12: ...of interest The type of electrodes plate pellet needle wire etc determine the type of lead wires needed to connect the electrodes to the input cable of the ECG amplifier and the preparation required...

Page 13: ...ding channel Turn the 3 way program switch to 1mV The signal on the recording channel should show a deflection of 1mVDC which is the nominal voltage used to calibrate ECG amplifiers If needed use the...

Page 14: ...10 World Precision Instruments...

Page 15: ...ct indirect special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused d...

Page 16: ...44 0 1462 424700 Fax 44 0 1462 424701 E mail wpiuk wpi europe com Germany Zossener Str 55 10961 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading...

Reviews: