WPI Biological Signals Amplifier Instruction Manual Download Page 11

Biological Signals Amplifier

World Precision Instruments 

7

Calibrating the Temperature Sensor to the TH Module

1.  In an insulated container, prepare 2–3L of water or fluid at about 20ºC. In a 

second insulated container prepare 2–3L of water at a higher temperature, 
about 40ºC.

2.  Put a precision thermometer (0.1ºC) through the lid of each insulated container 

so that the bulb of the thermometer is in the middle of the container. Leave an 
opening on the lid of each container for the temperature sensor.

3.  Place the temperature probe into the cooler fluid. Place the tip of the probe as 

close to the bulb of the thermometer as possible without touching it. 

4.  Turn the 3-way program switch to the 

MEAS

 position. Wait until the 

temperature reading on the TH module stabilizes. Then, read the temperature 
on the thermometer. Set the 3-digit 
display to the temperature of the 
thermometer by adjusting the 

LOW

 

potentiometer on the front of the TH 
module.

5.  Repeat Steps 3 and 4 for the warmer 

fluid in the other insulated container. 
Since the temperature of the fluid 
is warmer, adjust the 3-digit display 
to read the temperature of the 
warmer fluid by adjusting the 

HIGH

 

potentiometer. The TH module and 
sensor is now set and calibrated for a 
range of 20–45ºC.

Fig. 7—(Right) The TH temperature sensor.

Summary of Contents for Biological Signals Amplifier

Page 1: ...Biological Signals Amplifier for isolated heart Langendorff Neely and perfused organ systems World Precision Instruments INSTRUCTION MANUAL Serial No _____________________ 070212 www wpiinc com...

Page 2: ......

Page 3: ...prior written permission of World Precision Instruments Inc CONTENTS ABOUT THIS MANUAL 1 INTRODUCTION 1 Parts List 2 Unpacking 2 INSTRUMENT DESCRIPTION 3 Setting Up HG Pressure Amplifiers and Sensors...

Page 4: ...iv World Precision Instruments...

Page 5: ...y perfusate pressures temperature and ECG signals or surface potentials from isolated perfused heart Langendorff and working heart Neely preparations The system can also be used to measure perfusion p...

Page 6: ...at sales wpiinc com or 941 371 1003 to speak with one of our representatives regarding the type of electrodes needed to record the biopotentials of interest Unpacking Upon receipt of this instrument m...

Page 7: ...ls Amplifier World Precision Instruments 3 INSTRUMENT DESCRIPTION Fig 2 The front panel has the operation switches and indicators Fig 3 The back panel has the input and output sockets along with the m...

Page 8: ...cket of the HG1 pressure amplifier Connect the other end of the output cable to one of the analog inputs of the data acquisition unit 3 Before the transducer is connected to a perfusion cannula or cat...

Page 9: ...pressure module as described in the previous section 2 Turn the 4 way program switch on the front panel of HG amplifier to GND Set the BALANCE potentiometer to the middle position Turn on the amplifi...

Page 10: ...log inputs of the data acquisition unit 3 Turn the 3 way program switch to GND Place the temperature probe on or into the area where the temperature needs to be measured Turn on the amplifier unit The...

Page 11: ...ler fluid Place the tip of the probe as close to the bulb of the thermometer as possible without touching it 4 Turn the 3 way program switch to the MEAS position Wait until the temperature reading on...

Page 12: ...of interest The type of electrodes plate pellet needle wire etc determine the type of lead wires needed to connect the electrodes to the input cable of the ECG amplifier and the preparation required...

Page 13: ...ding channel Turn the 3 way program switch to 1mV The signal on the recording channel should show a deflection of 1mVDC which is the nominal voltage used to calibrate ECG amplifiers If needed use the...

Page 14: ...10 World Precision Instruments...

Page 15: ...ct indirect special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused d...

Page 16: ...44 0 1462 424700 Fax 44 0 1462 424701 E mail wpiuk wpi europe com Germany Zossener Str 55 10961 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading...

Reviews: