29
IPX8 (footswitch)
15.3 Device for intermittent operation: 60s ON, 10s OFF
15.4 Power-supply voltage:
~
220±22V 50Hz/60Hz 360mA (Max)
15.5 Fuses: 2×0.5AT 250V
15.6 Working frequency: 24kHz
~
29.5kHz
15.7 Flow: 25
~
100ml/m
15.8 Protection systems and tripping time of the APC:
No handpiece connected: 10ms
Cord interrupted: 10ms
Tips broken or not correctly tightened:
<
500ms
Protection by discharge to earth: 10ms
15.9 Alarm: Front display show the e (see point 7.3 and 14)
15.10 Operation environment:
a) Environment temperatuer: +10
℃~
+40
℃
b) Relative humidity: 30%
~
75%
c) Atmosphere pressure: 700hPa
~
1060hPa
d) Temperature in the water inlet of water-cooling equipment is not higher than
25
℃
15.11 Delivery and store environment: Store in the environment of humidity
under 80%, pressure around:50kPa
~
106kPa,temperature between -10
℃~
+50
℃
15.12 Handpiece cable: less than 100 sterilization cycles is highly recommended
15.13 Pump tube: less than 8 sterilization cycles is highly recommended