10
be connected to a supply mains with protective earth.
6.2 Initial installation
To ensure perfect operation of the equipment, it is installed by technical
personnel authorized by Woodpecker. The equipment will be installed in a
suitable and handy place for it to be used.
The technician must:
6.2.1 Install the device in a suitable place;
6.2.2 Explain the main aspects of correct installation to the user;
6.2.3 Fill in the installation form, including the purchaser’s data;
6.2.4 Send the installation form to Woodpecker to ensure traceability and
activation of the warranty.
Fig.5
6.3 Connection the accessories
The accessories listed as follow should be connected with the ULTRASURGERY:
6.3.1 Insert the silicone tube into the peristaltic pump, proceeding as follows:
a) Open the door(Fig.5—Ref.A)as far as it will go.
b) Position the tube in the impeller(Fig.5—Ref.B).
c) Close the door completely(Fig.5—Ref.C).
Danger: personnel injury.
The footswitch of the ULTRASURGERY must not be activated when the door of
the peristaltic pump open. (Fig.5—Ref.B).Moving parts could injure the operator.
6.3.2 Insert the rod for supporting the bag into the holes provided for it (Fig.6—
Ref.A);