8.5.5 Autoclave sterilizes the tips.
8.6 Autoclave sterilization of the wrench for tightening the tips
8.6.1 Clean the wrench;
8.6.2 Disinfect the wrench with a mild disinfectant solution having a neutral Ph 7
and dry it thoroughly;
8.6.3 Seal the wrench inside an individual disposable bag;
8.6.4 Autoclave sterilizes the wrench.
8.7 Autoclave sterilization of the peristaltic pump tube
8.7.1 Clean the peristaltic pump tube;
8.7.2 Disinfect with a mild disinfectant solution having a neutral pH 7 and dry it
thoroughly;
8.7.3 Seal the tube inside an individual disposable bag;
8.7.4 Autoclave sterilizes the tube.
8.8 Autoclave sterilization of the connection between cord and peristaltic pump
tube connection
8.8.1 Clean the connection between cord and peristaltic pump tube connection;
8.8.2 Disinfect with a mild disinfectant solution having a neutral pH and dry it
thoroughly;
8.8.3 Seal the connection inside an individual disposable bag;
8.8.4 Autoclave sterilizes the connection.
8.9 Autoclave sterilization of the handpiece support
8.9.1 Clean the handpiece support;
8.9.2 Disinfect with a mild disinfectant solution having a neutral pH 7 and dry it
thoroughly;
8.9.3 Seal the connection inside an individual disposable bag;
8.9.4 Autoclave sterilizes the connection.
9. Regular maintenance