Woodpecker ULTRASURGERY Manual Of Use And Maintenance Download Page 1

  

ULTRASURGERY 
MANUAL OF USE AND MAINTENANCE

Industrial design patent No.: CN 200830301694.8

Summary of Contents for ULTRASURGERY

Page 1: ...ULTRASURGERY MANUAL OF USE AND MAINTENANCE Industrial design patent No CN 200830301694 8...

Page 2: ...Connection the accessories 10 7 Controls 12 7 1 Description of the controls 12 7 2Description of the display and functions 13 7 3 Safety requirements during use 14 7 4 Protection systems and alarms 1...

Page 3: ...21 8 9 Autoclave sterilization of the handpiece support 21 9 Regular maintenance 21 10 Replacement of the fuses 22 11 Disposal procedures and precautions 23 12 Tips 23 13 Symbols 24 14 Troubleshootin...

Page 4: ...er with the equipment under any circumstances If any problems are encountered please contact a Woodpecker Service Centre Any attempts on the part of the user or any unauthorized personnel to tamper wi...

Page 5: ...enables osteotomy and osteoplasty techniques to be applied to in almost any anatomical situation This equipment can be used in the following fields a Oral surgery b Orthopedic surgery c Maxillofacial...

Page 6: ...he purpose for which it is intended see point 1 3 failure to comply with this requirement could lead to serious harm to the patient and or to the operator and or damage to failure of the equipment Dan...

Page 7: ...ch tips should definitely not be used It is necessary to instruct the patient to breathe though his nose during the treatment in order to avoid ingestion of the broken off fragment of the tip Danger D...

Page 8: ...e above information whenever you contact a Woodpecker Service Centre 2 2 Data plate of the device Each device has its own data plate Fig 1 on which technical specifications and serial number are indic...

Page 9: ...the failure parts This procedure ensures the function and reliability of all the parts 4 Delivery Avoid the excessive concussion shake cover in delivery Do not mix with the danger articles Avoid the s...

Page 10: ...dpiece 1 Fig 4 Ref F Plug of Footswitch 1 Fig 4 Ref G Footswitch 1 Fig 4 Ref H Tip Holder Marked on the packing list Fig 4 Ref I Tip Marked on the packing list Fig 4 Ref J Input of power supply cable...

Page 11: ...8 Fig 4...

Page 12: ...e protected from blows and from accidental sprays of water or other liquids Danger Do not install the device on or in the vicinity of sources of heat Install it such a way that there is an adequate ci...

Page 13: ...the installation form to Woodpecker to ensure traceability and activation of the warranty Fig 5 6 3 Connection the accessories The accessories listed as follow should be connected with the ULTRASURGER...

Page 14: ...holes provided for it Fig 6 Ref C 6 3 6 Insert the tube of ULTRASURGERY cord to the cord connector on the device Fig 4 Ref P 6 3 7 Put the handpiece on the support Fig 4 Ref O 6 3 8 Connect end of the...

Page 15: ...the controls This section illustrates the parts of the front panel of the ULTRASURGERY unit enabling the controls described in this manual to be located immediately 7 1 1Description of bone function F...

Page 16: ...ns There are three functions of bone root clean for this ULTRASURGERY 7 2 1 BONE function Fig 9 In bone function the power model is unavailable but we can adjust the water model four models as follows...

Page 17: ...cy vibrations and wear may very occasionally lead to breakage of the tip Tips of which the shape has been changed or which are otherwise damaged are liable to break during use Any such tips should def...

Page 18: ...it to detect the point where resonance of the tip is without any interference thus enabling optimum performance If this is not done contact with the part to be treated or with other surfaces before st...

Page 19: ...c casting Fig 8 and do not turn it while fastening the tip in place b Turn the wrench in a clockwise direction until the cultch engages till making clicking sound c The tip is now properly tightened i...

Page 20: ...6 4 Always make sure that any threaded parts and their contact surfaces are perfectly clean 7 6 5 If an tip becomes too worn the device will stop working 7 7 Settings permitted according to insert typ...

Page 21: ...conds and cannot be stopped e Once the cleaning cycle has been completed the device exits from the CLEAN function and returns to BONE function Fig 9 f On completion of the cleaning operations empty th...

Page 22: ...r viral infection always clean disinfect and sterilize the following components after each treatment a Handpiece Fig 3 b Tips Fig 4 Ref J c Torque wrench Fig 4 Ref C d Pump tube Fig 4 Ref P e Cord per...

Page 23: ...handpiece using a cloth moistened with a mild disinfectant solution having a neutral PH7 c Dry the electric contacts by blowing air onto them with the syringes d Seal the handpiece in an individual di...

Page 24: ...e an individual disposable bag 8 7 4 Autoclave sterilizes the tube 8 8 Autoclave sterilization of the connection between cord and peristaltic pump tube connection 8 8 1 Clean the connection between co...

Page 25: ...om the power mains Danger Check regularly that the power cable is intact if it is damaged replace it with an Woodpecker spare 10 Replacement of the fuses Danger Switch off the apparatus Always turn of...

Page 26: ...eat bone structures efficiently and effectively Sharp tips are used in osteotomy and osteoplasty when a fine and well defined cut in the bone structure concerned it required there are also tips with s...

Page 27: ...schneider s membrane or for lateralizing nerves In periodontology these tips are used to smooth the root surfaces 13 Symbols Trademark Use indoor only Alternating current Socker for the foot switch Te...

Page 28: ...nying documents Authorised Representative in the EUROPEAN COMMUNITY Appliance compliance WEEE directive Serial number Got the quality management system certification and CE certification issued by T V...

Page 29: ...he connector of the footswitch is not properly plugged into the socket Insert the footswitch connector properly The footswitch will not work Contact the nearest dealer or authorized Woodpecker service...

Page 30: ...unctioning of the tuning circuit Contact the nearest dealer or authorized Woodpecker service centre No liquid comes out of the tip during operation The tip is of the type with no through flow of liqui...

Page 31: ...at the door of the peristaltic pump is properly closed Insufficient power The tip is not correctly fitted to the handpiece the message WARN appears on the display Unscrew the tip and screw it back int...

Page 32: ...ischarge to earth 10ms 15 9 Alarm Front display show the e see point 7 3 and 14 15 10 Operation environment a Environment temperatuer 10 40 b Relative humidity 30 75 c Atmosphere pressure 700hPa 1060h...

Page 33: ...responsible for any irretrievable damage caused by the non professional person 17 Environmental protection There is not any harm factor in our product You can deal with it based on the local law 18 M...

Page 34: ...EN 60601 1 2006 EN 60601 1 2 2007 EN 60601 1 4 1996 EN 60601 1 6 2007 EN 61847 1998 EN 980 2008 EN 1041 2008 EN ISO 14971 2009 EN ISO 7405 2008 EN ISO 17664 2004 EN ISO 17665 1 2006 EN ISO 10993 1 20...

Page 35: ...32...

Page 36: ...33...

Page 37: ...ore being placed on the market all WOODPECKER equipment undergoes a thorough final check to ensure that it is are in proper working order 21 2 WOODPECKER guarantees its products purchased new from a W...

Page 38: ...nvironmental conditions in which the device is kept and stored do not comply with the requirements indicated in the chapter on technical specifications 21 5 Accidental damages due to transport incorre...

Page 39: ...ECKER produces will be charged on the basis of the actual cost of the materials and labour required fro such repairs 22 Statement This is to certify that all the functions of the equipment have been t...

Page 40: ...belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO LTD The industrial design inner structure etc have claimed for several patents by WOODPECKER any copy or fake product must take legal responsibilitie...

Page 41: ...ZMN WI 04 527 2 3Edition...