分类

The device is classified as IIa according to Rule 9 in Annex VIII of MDR (2017/745).
All active therapeutic devices intended to administer or exchange energy are classified
as class IIa.unless their characteristics are such that they may administer energy to or
exchange energy with the human body in a potentially hazardous way, taking account
of the nature, the density and site of application of the energy, in which case they are
classified as class IIb.
The contra-angle handpieces is a non-invasive device intended for short-term use, and
not direct in contact with human body especially the heart or central system.The way
of administration of energy or exchanging energy may indirectly injure the tissues in
the mouth when misused by operators or device under condition of malfunction. The
device with low risk because of it’s site of application.
The contra-angle handpieces not intended to control or monitor the performance of
active therapeutic class IIb devices, or not intended directly to influence the
performance of such devices.
The contra-angle handpieces not intended to emit ionizing radiation for therapeutic
purposes, including devices which control or monitor such devices, or which directly
influence their performance.
The contra-angle handpieces is not intended for controlling, monitoring or directly
influencing the performance of active implantable devices.

Based on the above analysis, the device is classed as IIa.

19.Device subcategory

MDA 0311 Active non-implantable dental devices.

Summary of Contents for CA161

Page 1: ...ece according to ISO 7785 2 and is use to hold and drive rotary instruments for the purposes of rotary processing intended for the endodontic treatment CA161 CA161 RR ISO 7785 2 Remark The figure belo...

Page 2: ...200 rpm 100 2500 rpm Output Torque Range 0 4 5 0 Ncm 0 6 4 2 Ncm Chuck type Push button Gear ratio 6 1 Rotation mode Continuous forward continuous reverse and reciprocating rotation Continuous forward...

Page 3: ...1 Type1 3 CA161 RR 1 Gear ratio 6 1 2 Output Speed Range 100 2500 rpm 3 Output Torque Range 0 6 4 2 N cm 4 Chuck type Push button 5 File shank attachment 2 35mm ISO1797 1 Type1 Table 1 3 Description o...

Page 4: ...2 3 Intended patient Patients who need root canal treatment 3 Operation Principle The Contra angle handpiece is a kind of geared angle handpiece and is use to hold and drive rotary instruments for th...

Page 5: ...ematopoietic system digestive system and endocrine system should be cautious to use this equipment 8 Pregnant women lactating women or women of childbearing age who have a plan of pregnancy should be...

Page 6: ...h force to treatment site 11 Please carry out regular maintenance and inspection 12 If the Contra angle Handpieces has not been used for a period of time before operation please maintain it as per mai...

Page 7: ...Output Speed Range 100 1200 rpm 100 2500 rpm Output Torque Range 0 4 5 0 Ncm 0 6 4 2 Ncm Chuck type Push button Gear ratio 6 1 Rotation mode Continuous forward continuous reverse and reciprocating ro...

Page 8: ...the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 8 Performance Item Technical requirement Appearance and Structure 1 The outer surface should be flat smooth and tidy wit...

Page 9: ...ece could be capable of being subjected to 250 cycles of sterilizing procedure 250 Preparation at the Point of Use The post operative process must be carried out immediately no later than 30 minutes a...

Page 10: ...nd be difficult to remove 10 40 Cleaning Regarding cleaning disinfection rinsing and drying it is to distinguish between manual and automated reprocessing methods Preference is to be given to automate...

Page 11: ...ector If necessary additional manual drying can be performed through lint free towel Insufflate cavities of instruments by using sterile compressed air If your washer disinfector does not have an auto...

Page 12: ...the device is obviously damaged smashed detached corroded or bent it must be discarded and not allowed to continue to be used If the accessories are found to be damaged please replace it beforeuse And...

Page 13: ...mum sterilization temperature 137 C Flash sterilization is not allowed on lumen instruments Storage Sterilized devices should be stored in a dry clean and dust free environment refer to label and inst...

Page 14: ...reference to previous and similar generations of the device 14 Statement Guilin Woodpecker Medical Instrument Co Ltd is not use of and will not use of the following materials to manufacture the Contr...

Page 15: ...ed 17 GMDN Code Code 38347 Definition A hand held mains electricity AC powered dental device that includes a chuck or collet for attaching a dental endpiece e g drill bit bur reamer saw blade polishin...

Page 16: ...or central system The way of administration of energy or exchanging energy may indirectly injure the tissues in the mouth when misused by operators or device under condition of malfunction The device...

Page 17: ...R Medical Instrument Co LTD File No ZMN YF BK10 01 001 001 1 Version A 20 Conformity assessment procedure Annex IX EU 2017 745 MDR 21 EU Representative Name MedNet EC Rep GmbH Add Borkstrasse 10 48163...

Reviews: