2).The air used for drying must be filtered by HEPA.
3)
The device should be dried in a clean area.
5)The drying temperature should not exceed 137 °C.
Maintenance
维护
1. Functional Test and Visual inspection
功能测试和目视检查
Visually inspect the cleanliness of the Handpiece. Perform Functional test
according to instructions of use. If there is still visible stain on the device after
cleaning, the entire cleaning process must be repeated.
Before packaging and sterilization, make sure that the contra-angle handpiece
has been maintained acc. to manufacturer's instruction.
If the device is obviously damaged, smashed, detached, corroded or bent, it
must be discarded and not allowed to continue to be used.
If the accessories are found to be damaged, please replace it beforeuse.
And the new accessories for replacement must be cleaned, disinfected
and dried.
2. Use lubricant to lubricate the handpiece and dried it prior to sterilization.
Aim the nozzle of lubricant bottle to the air hole at the end of the contra-angle
handpice to inject oil for 1-2 seconds.
Packaging
The contra-angle handpiece should be quickly packaged in a medical
sterilization bag (or special holder, sterile box).
Precautions
1
)
Only use a legally marketed sterilization pouch;
2
)
)
The package should withstand high temperature of 137 °C and has
sufficient steam permeability;
3)
The packaging environment and related tools must be cleaned regularly to