Preparation for
reprocessing

The products must be reprocessed in a disassembled state.

产品必须在拆卸状

态下进行再加工。

Pre-cleaning

预清

Tools: tray, soft brush, clean and dry soft cloth
Perform manual pre-cleaning until the handpiece is visually clean. Rinse the
chuck of bur with running water for at least 10 seconds. Clean the surface with
a soft bristle brush.
Note: The water temperature should not exceed 40

during the washing

stage, otherwise the protein will solidify and be difficult to remove.

工具:托盘、软刷、清洁干燥的软布

执行手动预清洁,直到机头目测清洁。 用流水冲洗车针夹头至少

10

秒钟。

用软毛刷清洁表面。

注意:洗涤阶段水温不宜超过

40

℃,否则蛋白质会凝固,难以去除。

Cleaning

Regarding cleaning/disinfection, rinsing and drying, it is to distinguish
between manual and automated reprocessing methods. Preference is
to be given to automated reprocessing methods, especially due to the
better standardizing potential and industrial safety.

关于清洗

/

消毒、漂洗和干

燥,是区分手动和自动再处理方法。 优先考虑自动再处理方法,特别是

由于

更好地标准化潜在和工业安全。

Automatic cleaning

自动清洗

The washer-disinfector should meet the requirements of the ISO 15883. Place
the contra-angle handpiece in the washer-disinfector carefully.
Ensure that contra-angle handpieces can not move freely in the
washer-disinfector.

The contra-angle handpiece are not permitted to contact with each other. Start

the program:

4min pre-washing with cold water(<40

);

Emptying

5 min washing with a mild alkaline cleaner at 55

℃;

Summary of Contents for CA161

Page 1: ...ece according to ISO 7785 2 and is use to hold and drive rotary instruments for the purposes of rotary processing intended for the endodontic treatment CA161 CA161 RR ISO 7785 2 Remark The figure belo...

Page 2: ...200 rpm 100 2500 rpm Output Torque Range 0 4 5 0 Ncm 0 6 4 2 Ncm Chuck type Push button Gear ratio 6 1 Rotation mode Continuous forward continuous reverse and reciprocating rotation Continuous forward...

Page 3: ...1 Type1 3 CA161 RR 1 Gear ratio 6 1 2 Output Speed Range 100 2500 rpm 3 Output Torque Range 0 6 4 2 N cm 4 Chuck type Push button 5 File shank attachment 2 35mm ISO1797 1 Type1 Table 1 3 Description o...

Page 4: ...2 3 Intended patient Patients who need root canal treatment 3 Operation Principle The Contra angle handpiece is a kind of geared angle handpiece and is use to hold and drive rotary instruments for th...

Page 5: ...ematopoietic system digestive system and endocrine system should be cautious to use this equipment 8 Pregnant women lactating women or women of childbearing age who have a plan of pregnancy should be...

Page 6: ...h force to treatment site 11 Please carry out regular maintenance and inspection 12 If the Contra angle Handpieces has not been used for a period of time before operation please maintain it as per mai...

Page 7: ...Output Speed Range 100 1200 rpm 100 2500 rpm Output Torque Range 0 4 5 0 Ncm 0 6 4 2 Ncm Chuck type Push button Gear ratio 6 1 Rotation mode Continuous forward continuous reverse and reciprocating ro...

Page 8: ...the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 8 Performance Item Technical requirement Appearance and Structure 1 The outer surface should be flat smooth and tidy wit...

Page 9: ...ece could be capable of being subjected to 250 cycles of sterilizing procedure 250 Preparation at the Point of Use The post operative process must be carried out immediately no later than 30 minutes a...

Page 10: ...nd be difficult to remove 10 40 Cleaning Regarding cleaning disinfection rinsing and drying it is to distinguish between manual and automated reprocessing methods Preference is to be given to automate...

Page 11: ...ector If necessary additional manual drying can be performed through lint free towel Insufflate cavities of instruments by using sterile compressed air If your washer disinfector does not have an auto...

Page 12: ...the device is obviously damaged smashed detached corroded or bent it must be discarded and not allowed to continue to be used If the accessories are found to be damaged please replace it beforeuse And...

Page 13: ...mum sterilization temperature 137 C Flash sterilization is not allowed on lumen instruments Storage Sterilized devices should be stored in a dry clean and dust free environment refer to label and inst...

Page 14: ...reference to previous and similar generations of the device 14 Statement Guilin Woodpecker Medical Instrument Co Ltd is not use of and will not use of the following materials to manufacture the Contr...

Page 15: ...ed 17 GMDN Code Code 38347 Definition A hand held mains electricity AC powered dental device that includes a chuck or collet for attaching a dental endpiece e g drill bit bur reamer saw blade polishin...

Page 16: ...or central system The way of administration of energy or exchanging energy may indirectly injure the tissues in the mouth when misused by operators or device under condition of malfunction The device...

Page 17: ...R Medical Instrument Co LTD File No ZMN YF BK10 01 001 001 1 Version A 20 Conformity assessment procedure Annex IX EU 2017 745 MDR 21 EU Representative Name MedNet EC Rep GmbH Add Borkstrasse 10 48163...

Reviews: