产品描述

Product Description

1. Device Composition

设备组成

Woodpecker Contra-angle handpiece (model: CA161

CA161-RR) is a kind of

geared angle handpiece according to ISO 7785-2 and is use to hold and drive rotary
instruments for the purposes of rotary processing, intended for the endodontic
treatment.

啄木鸟弯手机

(

型号

: CA161

CA161-RR)

是一种符合

ISO 7785-2

标准的齿轮

角手机,用于固定和驱动旋转器械以进行旋转加工,用于牙髓治疗。

Remark: The figure below is schematic diagram.Differences between different models
in the table 1-1,The components in the table 1-2 and attachments in the table 1-3 are
sold together with the device.

备注:下图为示意图。表

1-1

中不同型号的区别,表

1-2

中的组件和表

1-3

中的附件与设备一起销售。

Figure 1 Structure of CA161

Summary of Contents for CA161

Page 1: ...ece according to ISO 7785 2 and is use to hold and drive rotary instruments for the purposes of rotary processing intended for the endodontic treatment CA161 CA161 RR ISO 7785 2 Remark The figure belo...

Page 2: ...200 rpm 100 2500 rpm Output Torque Range 0 4 5 0 Ncm 0 6 4 2 Ncm Chuck type Push button Gear ratio 6 1 Rotation mode Continuous forward continuous reverse and reciprocating rotation Continuous forward...

Page 3: ...1 Type1 3 CA161 RR 1 Gear ratio 6 1 2 Output Speed Range 100 2500 rpm 3 Output Torque Range 0 6 4 2 N cm 4 Chuck type Push button 5 File shank attachment 2 35mm ISO1797 1 Type1 Table 1 3 Description o...

Page 4: ...2 3 Intended patient Patients who need root canal treatment 3 Operation Principle The Contra angle handpiece is a kind of geared angle handpiece and is use to hold and drive rotary instruments for th...

Page 5: ...ematopoietic system digestive system and endocrine system should be cautious to use this equipment 8 Pregnant women lactating women or women of childbearing age who have a plan of pregnancy should be...

Page 6: ...h force to treatment site 11 Please carry out regular maintenance and inspection 12 If the Contra angle Handpieces has not been used for a period of time before operation please maintain it as per mai...

Page 7: ...Output Speed Range 100 1200 rpm 100 2500 rpm Output Torque Range 0 4 5 0 Ncm 0 6 4 2 Ncm Chuck type Push button Gear ratio 6 1 Rotation mode Continuous forward continuous reverse and reciprocating ro...

Page 8: ...the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 8 Performance Item Technical requirement Appearance and Structure 1 The outer surface should be flat smooth and tidy wit...

Page 9: ...ece could be capable of being subjected to 250 cycles of sterilizing procedure 250 Preparation at the Point of Use The post operative process must be carried out immediately no later than 30 minutes a...

Page 10: ...nd be difficult to remove 10 40 Cleaning Regarding cleaning disinfection rinsing and drying it is to distinguish between manual and automated reprocessing methods Preference is to be given to automate...

Page 11: ...ector If necessary additional manual drying can be performed through lint free towel Insufflate cavities of instruments by using sterile compressed air If your washer disinfector does not have an auto...

Page 12: ...the device is obviously damaged smashed detached corroded or bent it must be discarded and not allowed to continue to be used If the accessories are found to be damaged please replace it beforeuse And...

Page 13: ...mum sterilization temperature 137 C Flash sterilization is not allowed on lumen instruments Storage Sterilized devices should be stored in a dry clean and dust free environment refer to label and inst...

Page 14: ...reference to previous and similar generations of the device 14 Statement Guilin Woodpecker Medical Instrument Co Ltd is not use of and will not use of the following materials to manufacture the Contr...

Page 15: ...ed 17 GMDN Code Code 38347 Definition A hand held mains electricity AC powered dental device that includes a chuck or collet for attaching a dental endpiece e g drill bit bur reamer saw blade polishin...

Page 16: ...or central system The way of administration of energy or exchanging energy may indirectly injure the tissues in the mouth when misused by operators or device under condition of malfunction The device...