direction.
④
Sterilize it in sterilizer after each treatment.
⑤
The torque wrench must be cooled naturally after sterilization to avoid
scald when using next time.
⑥
Keep the torque wrench in a cool, dry and ventilated place and be clean.
c
) Precaution
The following sterilizing methods are forbidden.
①
Braise in liquor;
②
Dip in iodine,alcohol or glutaraldehyde;
③
Torrefy in oven or microwave oven.
Notice:we are not responsible for any damage of the torque wrench directly
or indirectly made by any way in the above items.
2.2 Endo function
a) Usage process
①
Fix endochuck to handpiece by endo wrench.
②
Unscrew the screw cap on the endochuck.
③
Put the ultrasonic file into the hole in the front of endochuck.
④
Screw down the screw cap with endo wrench to tight up the ultrasonic file.
⑤
Press option key, turn to endo function.
⑥
When ultrasonic scaler turns into endo function, only the first lead light is
on and the power is at 1st grade. Put the ultrasonic file into the patient's root
canal slowly, step on the foot switch, then make endo treatment. During the
treatment, turn up the power gradually according to the needs.
b) Notice
①
When fixing endochuck, it must be screwed down.
②
The screw on the endochuck must be screwed down.
③
Don't press it too much when the ultrasonic file in root canal.
④
Don't start up the foot switch until the ultrasonic file is in root canal.
⑤
The power range is supposed from 1st to 5th grades.
9