33
Veterinary Lactate Meter
Troubleshooting
This section details the significant display screen messages and error codes you
may encounter when using the L-Pet Veterinary Lactate Meter.
If you have further questions after reviewing these messages, call your
authorised dealer.
hWhat To Do
Message
Problem
What To Do
“Code- - - -"
“Code- - - -"
The monitor has not
been coded with a
code card.
The code card has
been damaged.
Insert the code card that is
included in the test strip package
into the code card port. Then,
insert a test strip into the test strip
holder.
Insert a new code card from a new
package of test strips, follow the
procedures on page 13, ‘Coding
the Meter’. If the problem persists,
call the authorised dealer.