Veterinary Lactate Meter
Step 4: Apply a drop of control solution to the yellow reaction zone in the middle
top of the test strip.
Step 5: The screen will show timing bars “----” that flash and then gradually
diminish for countdown.
Step 6: After the timing bars disappear (approx. 45 seconds), the screen will
show the test result. Compare the reading on the screen to the range
indicated on the test strip label.
Step 7: Remove the test strip and discard it into clinical waste.
Step 8: The screen will display the blood test mode.
The test result of control solution will not be stored in memory.
NOTE
:
If the test result is not within the expected range, repeat the test until the result
reports within the range. If the results continue to report outside of the expected
range, please contact the authorised dealer.