Veterinary Lactate Meter

Step 4: Apply a drop of control solution to the yellow reaction zone in the middle 

top of the test strip.

Step 5: The screen will show timing bars “----” that flash and then gradually 

diminish for countdown.

Step 6: After the timing bars disappear (approx. 45 seconds), the screen will 

show the test result. Compare the reading on the screen to the range 
indicated on the test strip label. 

Step 7: Remove the test strip and discard it into clinical waste.
Step 8: The screen will display the blood test mode.
The test result of control solution will not be stored in memory.

NOTE

If the test result is not within the expected range, repeat the test until the result 
reports within the range. If the results continue to report outside of the expected 
range, please contact the authorised dealer.

Summary of Contents for L-Pet

Page 1: ......

Page 2: ...Test Strip 7 Using the Meter for the First Time 9 Setting the Clock 9 Turning ON OFF Beep Sound 10 Changing Unit of Measurement 11 Coding the Meter 13 Checking the System 16 Control Solution Method 1...

Page 3: ...Veterinary Lactate Meter 2 Using the Meter Memory 27 Setting Time After First Use 28 Meter Maintenance 30 Troubleshooting 33 Product Warranty 36 Specifications 37 Symbols Used in this Manual 39...

Page 4: ...ing individual metabolism and oxygen debt Lactate results can be used to detect local or systemic hypoperfusion and monitor response to therapy There is also a strong correlation between lactate conce...

Page 5: ...r Lancing Device L Pet Blood Lactate Test Strips 5 Lancets 28G Monitor Checker Wallet 3V Lithium Coin Cell Battery CR2032 Code Card User Guide Log Book In addition to the items above the L Pet Lactate...

Page 6: ...t Display Battery Compartment Enter Button Shift Button Test Strip Holder FRONT BACK Veterinary Lactate Meter Woodley Equipment Company Ltd St John Street Horwich Bolton BL6 7NY UK Tel 44 0 1204 66903...

Page 7: ...ocated on the lower front side of the meter Battery Compartment Holds one 3V Lithium Coin Cell Battery to power the system Located on the back of the meter Monitor Checker Used to confirm that the met...

Page 8: ...trip Reaction Zone The area where the blood sample or control solution is applied IMPORTANT L Pet Blood Lactate Test Strips are contained in a moisture proof and light protected bottle The test strips...

Page 9: ...bottle DO NOT leave test strips outside the bottle when not in use When performing a blood test insert the test strip with contact points facing up and towards the meter then apply the blood sample Fo...

Page 10: ...ss and hold SHIFT for quick searching of the years The meter provides calendar years from 2005 to 2099 Step 3 The pre set Month will flash on the screen Press SHIFT to scroll through the months and pr...

Page 11: ...en Press SHIFT to scroll through the minutes and press ENTER to confirm Press and hold SHIFT for quick searching of the minute Step 7 Press the two buttons on the meter simultaneously to return to the...

Page 12: ...Press ENTER to enter setup Press SHIFT to turn the beep sound ON OFF The speaker with a wave ahead of it means ON the speaker without the wave means OFF Step 7 Press the two buttons on the meter simul...

Page 13: ...SHIFT button to skip this selection Step 6 A symbol of a speaker will flash on the screen Press SHIFT button to skip this selection Step 7 A symbol of a clock will flash on the screen Press SHIFT to...

Page 14: ...r on the package of test strips before each use of the meter The code needs to be set only once for each package of test strips The meter will memorise the code until it is changed If the code number...

Page 15: ...test strips Step 2 Verify that the code number on the code card matches the code number on the test strip package Step 3 Insert the code card with the code number facing up firmly and completely into...

Page 16: ...e number displayed on the screen should match the code number on the grey code card and on the front of the test strips package NOTE Recoding needs to be performed when opening a new package of test s...

Page 17: ...the meter and test strips are working together properly Monitor Checker Method It is recommended that the L Pet Veterinary Lactate Meter is checked using this method when the meter is new and when th...

Page 18: ...en self testing is complete Remove the Monitor Checker from the test strip holder and the screen will then display Ctrl Step 3 Press ENTER to exit checking the meter NOTE If the screen displays nt OK...

Page 19: ...the test strip are functioning properly When to perform a control solution test When the meter is used for the first time Whenever there is doubt that the meter or the test strips are not working prop...

Page 20: ...the test strip label DO NOT use the system Repeat the test until the result reports within the expected range If the results continue to report outside the expected range contact the authorised deale...

Page 21: ...le Step 2 The screen will display Ctrl and then insert an L Pet Blood Lactate Test Strip Step 3 The screen will flash a symbol of a liquid drop Squeeze a drop of control solution onto a clean dry non...

Page 22: ...show the test result Compare the reading on the screen to the range indicated on the test strip label Step 7 Remove the test strip and discard it into clinical waste Step 8 The screen will display the...

Page 23: ...lancing device cap 2 Place the lancet into the lancing device 3 Twist and pull the protective lancet cover to expose the sterile needle tip 4 Screw on the lancing device cap Select the skin penetrati...

Page 24: ...or if the CODE number on the screen does not match the CODE number on the test strip package refer to page 13 for details of re coding the meter Step 2 The meter will self test the environment tempera...

Page 25: ...est result Step 3 Obtain a drop of whole blood and apply to the absorbent area at the curved edge of the test strip NOTE DO NOT touch the target area of the test strip DO NOT smear the blood drop onto...

Page 26: ...spose in clinical waste The LCD screen on the meter will show the code number and a flashing arrow next to the test strip icon indicating the meter is ready for another test Step 7 If more tests are t...

Page 27: ...tate Meter 26 Understanding the Test Result Normal Range 2 5 mmol L 23 mg dL Mild Lactate Shock 2 5 5 0 mmol L 23 45 mg dL Moderate Lactate Shock 5 0 7 0 mmol L 45 63 mg dL Severe Lactate Shock 7 0 mm...

Page 28: ...sult is added When the test results are recalled from memory the most recent result is displayed first To view the results from memory Step 1 Press any key to turn on the meter and press two buttons o...

Page 29: ...his selection Step 4 A symbol of a cable will flash on the screen Press SHIFT to skip this selection Step 5 A symbol of a speaker will flash on the screen Press SHIFT to skip this selection Step 6 A s...

Page 30: ...SHIFT to scroll through the days and press ENTER to confirm Press and hold SHIFT for a quick search of the day Step10 The Hour will flash on the screen The meter uses AM PM time clock Press SHIFT to...

Page 31: ...s dropping the meter could damage the electronics DO NOT try to clean the test strip holder DO NOT contaminate the strip holder with blood or control solution Storage The meter is designed to be used...

Page 32: ...indicates the battery is low and should be replaced as soon as possible NOTE To save battery power the meter will turn off automatically after one minute of inactivity All results stored in memory wi...

Page 33: ...time after replacing the battery See page 28 Setting Time After First Use Cleaning the Meter To clean the outside of the meter use a lint free cloth dampened with soapy water or alcohol NOTE DO NOT ge...

Page 34: ...your authorised dealer hWhat To Do Message Problem What To Do Code Code The monitor has not been coded with a code card The code card has been damaged Insert the code card that is included in the tes...

Page 35: ...the meter s operating range of 10 C to 40 C 50 F to 104 F Monitor Checker result fails Used or defective test strip Replace the battery The meter needs to be moved to an area that is within the meter...

Page 36: ...I Err The blood lactate result is lower than 0 7 mmol L 6mg dL The blood lactate result is higher than 22 2 mmol L 200 mg dL The meter is damaged Repeat the test to confirm the test result Repeat the...

Page 37: ...ted to repair or replacement and in no event shall Woodley Equipment Company Ltd be liable for any collateral or consequential damages or loss Instruments subjected to misuse abuse neglect unauthorise...

Page 38: ...lume Minimum 3 L Reading Time 45 seconds Hematocrit Range 35 50 Memory 300 sets with date time Operating Temperature 10 C to 40 C 50 F to 104 F Relative Humidity Less than 85 Storage Operating and Tra...

Page 39: ...ctate Meter 38 Electromagnetic Compatibility This equipment complies with the EMC requirement of EN 60601 1 2 Classification according IPXO not evaluated as AP APG equipment to IEC EN 60601 1 continuo...

Page 40: ...39 Veterinary Lactate Meter Symbols Used in this Manual Use by date Batch code Temperature limit Consult instructions for use Caution Keep away from sunlight Keep dry Do not re use Manufacturer...

Page 41: ......

Reviews: