16
Prior to testing, clean the testing site with an alcohol swab or by washing it with warm,
soapy water. Dry the testing site thoroughly.
Hold the lancing device against the paw/ear. Push to release button to prick the site. A click
should be heard as the lancing device activates. Avoid smearing the drop of blood.
Note:
Don’t use an infection swab containing iodine
to clean the puncture site as this can
cause inaccurate results.
Disposal of the Lancet
Unscrew the lancing device cover. Place the safety tab back onto the lancet needle. Pull the
used lancet off. Place the lancing device cover back on the lancing device.
Note:
For veterinary use only.
Safety Lancets
1.
Carefully rotate and pull off the protective cap
2.
After cleaning the skin, hold the lancet firmly against the puncture site
3.
Press the lancet against the puncture site. Discard the lancet in an appropriate
sharps container
Test Processing
Ensure the meter is set up properly as described in Sections 4 and 5. Turn the meter on.
Ensure the code chip is inserted. Compare the number shown on the display with the code
number printed on the box label or foil pouch. The test strip icon flashes to indicate that it is
ready for the strip to be inserted.
Check the sample type on the meter is the same as the sample type collected. If not, set the
correct sample type as described in Section 5.
Testing
Insert a test strip into the Test Strip Holder in the same direction as the arrows indicate on
the strip. Ensure that the test strip is inserted all the way to the end of the Test Strip Holder.