User’s Manual – PGT
®
130.DT
Part No.: 7100.PGT130.DT
Page 7 / 21
V12/2020
Parameter
Funktion
Language
Device language selection
English
, German, Spanish, French
Voltage
Measure voltage
30V, 50V,
100V
Measure mode
Measure modes
AND - Shoe and wrist required
OR
- Shoe or wrist measurement
WRIST - Only wrist measurement required
SHOE - Only shoe measurement required
Buzzer
Internal buzzer
ON
, OFF
Low Limit
Low limit for shoe measurement (100k
Ω
) and wrist strap system
measurement (750k
Ω
)
ON
, OFF
High Limit Shoe
High limit for shoe measurement
20 M
Ω
, 35M
Ω
, 70 M
Ω
,
100 M
Ω
High Limit Wrist strap
High limit for wrist strap / smock measurement
35 M
Ω
, 100M
Ω
, 1000 M
Ω
*
(
*
100V Measure voltage required)
Hands-free mode
Measurement without hands (shoe to shoe)
ON,
OFF
Relay pulse length
Relay pulse length
1s
, 2s, 3s
Internal RFID
Internal RFID reader
ON
, OFF
Calibration
Measurement display for device calibration
Default settings are in bold
Initial Operation via Network
The PGT
®
130.DT is ready within 50 seconds after plugging-in the power supply. If the Ethernet
cable is connected, the device tries to configure its network by DHCP. Therefore a DHCP server
should be available within the network. A label indicating the assigned IP Address will be printed on
system start-up if the label printer is connected. The IP-Address is displayed temporarily on the LCD
display as well.
System clock can be set by DHCP as well (042 timeserver).
If Internet access is available, the system clock will synchronized with a NTP timeserver.
The configured DHCP hostname will be sent to the DHCP server if available.
If DHCP configuration fails the default IP-Address
192.168.1.1
will be used.
It is possible to enter a different IP-Address manually.
Entering a fixed IP-Address will deactivate DHCP.
To configure the Data Terminal enter the assigned IP-Address into your browser's URL-field.
The first access opens the configuration page without asking for a password.
Adjust and safe the settings to continue.