User’s Manual – PGT
®
130.DT
Part No.: 7100.PGT130.DT
Page 6 / 21
V12/2020
Installation
1. Connect the footwear electrode to the red and black socket.
2. If you wish to use external input device connect it to the USB2 socket.
3. Connect the external Temperature/Humidity sensor via supplied extension cable to the
device if applicable.
4. If you’re using printer functionality connect the label printer to a free USB port.
5. If applicable, connect the network cable to the LAN socket.
6. Connect the supplied power supply to the PWR socket.
Use only the original power supply which was included in the scope of delivery.
Rebooting is required to activate new connected USB devices.
Initial Operation - Local
If you operate the device in stand-alone you can configure it in front via display. The symbols for
wrist and shoe are touch sensitive.
Info:
This option can be disabled via web interface settings [Device settings – Config buttons
]
Press the left button for at least 3 seconds
Keep the left button pressed and tap the right button at least 4 times back-to-back
The device switches to configuration mode. The configuration mode does time-out automatically
after 5 minutes without operation, settings will stored.
Browse through the parameter menu with the left button. After the last parameter the
configuration mode does exit and the settings are stored.
Switch between the available settings with the right button.