background image

Wöhler FA 320

 

2. Specifications 

2.1 Application 

The Wöhler FA 320 Fan Anemometer is a handy, handheld instrument designed 
to measure the flow velocity, temperature and relative humidity of air. It can be 
used for example for the measuring of ventilation plants, in housing spaces etc. 
Because of the integrated average determination of eight measuring points it will 
be possible to measure even large outlets. In this case the air volume flow will be 
calculated automatically. 
The user can select between different units (°C, °F, %rH, m/s, ft/m, m³/m, cfm, 
cm, inch). 

2.2 Measured 

values 

Flow velocity: 
Flow velocity range: 

0,3 … 35 m/s 

Resolution:  

0,1 

m/s 

Accuracy: 

 

± 5 % o.m. 

 
Air volume: 
Air volume range: 

0 … 99.999 m³/s 

Resolution: 

 

0,1 (0 … 9.999,9 m³/s) 

 

 

 

1 (10.000 … 99.999 m³/s) 

Accuracy: 

 

± 5 % o.m. 

 
Temperature: 
Temperature range: 

-20 °C … 60 °C 

Resolution:  

0,1 

°C 

Accuracy:   ± 

°C 

Summary of Contents for 22107

Page 1: ...er Art Nr 22107 Contents 1 Preface 12 2 Specifications 13 3 Operations 14 4 Declaration of Confirmity 19 5 Guarantee and Service 20 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1...

Page 2: ...fore com missioning the unit In principle the W hler FA 320 Fan Anemometer should only be used by skilled personnel for its intended purpose and within the specified range of data Any liability or war...

Page 3: ...age determination of eight measuring points it will be possible to measure even large outlets In this case the air volume flow will be calculated automatically The user can select between different un...

Page 4: ...al Data Dimensions vane 170 x 77 x 40 mm Dimensions meter 175 x 70 x 33 mm Weight vane 150 g Weight meter 185 g Power 4 x 1 5 V battery AAA 3 Operation 3 1 Important Reminder Rotate the cover before m...

Page 5: ...oint wet bulb relative humidity velocity 3 2 3 ENTER Press ENTER to confirm the setting and calibration 3 2 4 Mx Mn UP Press to view MAX MIN value Press to select the value of each digital cycle 3 2 5...

Page 6: ...press the POWER key for more than two seconds to enter the unit selection mode Press or to select the unit and then press ENTER to save C RH m s m m cm or F RH ft m cfm inch 3 3 4 Velocity with averag...

Page 7: ...setting and enter the measuring mode see 3 4 2 Diameter While in default outlet setting mode length press the ENTER key for over two seconds to choose the diameter setting Press the HOLD key to selec...

Page 8: ...eps when the 60 seconds are over After the time is up the 60 seconds average volume will be indicated in the dis play If there is more than one outlet to be measured press the ENTER key again to repea...

Page 9: ...ntial protection requirements of Council directive 89 336 EC 92 31 EEC and 93 68 EEC on the approximation of the laws of the Member States relating to electromagnetic compability The meter conforms to...

Page 10: ...centers listed below You can obtain immediate help from our engineers by telephone 5 3 Information on disposal You may hand in any defective batteries taken out of the unit to our company as well as...

Reviews: