Page 2
|
Table of Contents
1
System Integration
The AMC100 and the AMC300 can be integrated with external systems or devices by
-
combining it with third party hardware
-
establishing incoming and outgoing trigger connections (/IO upgrade required)
-
controlling it with individual software interfaces
1.1
Connecting to Third Party Hardware
CAUTION
General hazard!
Inadequate hardware connections may cause injury and are likely to damage the device or
interfere with an appropriate functioning.
o
Always contact attocube for technical support, before combining the device with
third party hardware.
o
Do carefully observe the information in this section when combining the device
with third party hardware.
NOTE
attocube is not liable for any damages resulting from an unauthorized combination of the
device with third party hardware. Unauthorized combination with third party hardware is
not covered by attocube's warranty.
1.1.1
Cabling Restrictions
For optimal performance, obey the following combination restrictions:
-
Do not to connect cabling with a wire resistance > 5 Ω.
-
Use EMV housings as enclosure for the D-sub connectors.
-
Use extra shielded twisted pair wires for the piezo voltage supply.
-
Do not connect any cable > 5 m.
1.1.2
Pin Assignments
The pins of the device’s positioner control cables can be found in the AMC100 or AMC300 User
Manual.