4.7
Program configuration
4.7.1
Defining a test cycle
In the “Test program” menu, you can open the “Program configurator…” and define a suitable test cycle
with the parameters for your test sample.
Here, you have the option of selecting one of the following three programs.
Automatic test
(The ChronoProof PRO automatically selects the suitable pressure)
Note:
Two pressures are always used for
testing: negative pressure (vacuum)
and excess pressure.
Test with 1 pressure
(adjustable from -0.5 to max. 3.0 bar)
Test with 2 pressures
(adjustable from -0.5 to max. 3.0 bar)
Notes:
-
You can set whether to use one or two pressures for testing.
-
The pressures are adjustable in steps of 0.1bar.
-
Both measurement pressures can be set as vacuum or pressure.
Important!
-
The first test pressure must always be lower than the second test pressure (however, incorrect entries
are not permitted by the program)