Test specification for immunity to RF wireless communications devices
Test frequency (MHz)
Modulation
Immunity test level (V/m)
385
Pulse Modulation : 18 Hz
27
750
FM + 5 Hz deviation : 1 kHz sine
28
710
– 745 – 780
Pulse Modulation : 217 Hz
9
810
– 870 – 930
Pulse Modulation : 18 Hz
28
1720
– 1845 – 1970
Pulse Modulation : 217 Hz
28
2450
Pulse Modulation : 217 Hz
28
5240
– 5500 - 5785
Pulse Modulation : 217 Hz
9
3.2.4. Equipotentiality
Under the head-half of the bed base you will find an equipotentiality socket
, identified by the label
,
enabling you to connect any electromedical devices. The leads of these devices must pass through the head
end and not on the side.
Label
Equipotentiality