Winncare IPx4 User Manual Download Page 6

(A) 

(B) 

2

 

INSTALLATION AND OPERATION 

2.1

 

Mounting of the DPS

 

Mounting the DPS 

The DPS is mounted the same way as other Winncare 
Nordic spreader bar models with strap.  

1)

 

Lift the spreader bar. Hold the strap and depress 
the safety spring-plate on the hook. 

2)

 

Attach the hook onto the attachment point on 
the spreader bar and release the safety spring 
plate. 

3)

 

Ensure that the hook is safely locked onto the 
spreader bar attachment point. 

4)

 

Connect the DC cable to the Luna. The socket is 
located on the side opposite the hand control 
socket. 

The DPS is ready for use if the instruction has been 
followed correctly. 

Detaching the DPS 

1)

 

Detach the DC cable 

2)

 

Press in the spring-plate on the hook. 

3)

 

Tilt the hook to detach the spreader bar. 

 

2.2

 

Operating the DPS 

Operating the DPS 

1)

 

Connect the hand control to the Luna ceiling 
hoist. 

 

2)

 

Press one of the horizontal buttons (A or B) to 
activate the DPS.  

Button A moves the DPS to the sitting position and button 
B to lying position as seen in the pictogram. 

 

 

 

Summary of Contents for IPx4

Page 1: ...Please read this user manual carefully before using the product This user manual is intended as a guide for caregivers and other users as well as technical personnel during servicing and maintenance...

Page 2: ...2...

Page 3: ...___________ 7 3 2 Bed to chair__________________________________________ 7 3 3 Chair to bed _________________________________________ 7 4 MAINTENANCE______________________________________________ 8 4...

Page 4: ...ers See 4 1 NOTE Do not use DPS for purposes other than originally intended 1 2 Safety instructions Important Warning Read user manual and safety instructions before using the product NEVER use the DP...

Page 5: ...nt Contents of the box 1 The DPS 2 User manual 3 Label for hand control 1 6 Transport If the DPS is to be transported between sites it is recommended to use the original packaging In case of return sh...

Page 6: ...o the spreader bar attachment point 4 Connect the DC cable to the Luna The socket is located on the side opposite the hand control socket The DPS is ready for use if the instruction has been followed...

Page 7: ...hair When lifting a patient from a bed the DPS should be positioned in the lying position For a smooth transfer the caregiver can use the bed s adjustment functions if available Position the patient i...

Page 8: ...luding ceiling hoist It is the legal owner s duty to ensure that these service checks are carried out on the product The following items is to be inspected during the annual check Bolts o Check for lo...

Page 9: ...erly connected Attach DC cable as seen in 2 1 Check that the DC cable going into the actuator is connected Defective hand control Check if up down buttons on the hand control works If not try a differ...

Page 10: ...10 6 2 Dimensions Measurements All dimensions in mm Lying position Sitting position A Height 550 750 B Width 525 170 C Track to sling attachment point 864 1070 C A B...

Page 11: ...y damage the environment if not handled in an environmentally responsible way 7 2 Quality Winncare Nordic is certified according to the quality standard ISO 9001 and ISO 13485 These certifications mea...

Page 12: ...nufacturer Winncare Nordic ApS Taarnborgvej 12 C 4220 Korsoer Denmark Phone 45 70 27 37 20 Fax 45 70 27 37 19 E mail info Winncare dk www Winncare dk 93 42 E F CLASS I EN ISO 10535 EN 60601 1 2 910 00...

Reviews: