PURPOSE AND SAFETY

 

The tilting spreader bar is an add-on for the Luna IPx4 2 channel model. The tilting spreader bar 
works as a dynamic positioning system (DPS) and will be referred to as 

DPS

 in this user guide. 

The DPS is powered, meaning that it is operated by the Luna hand control. This allows for 
powered patient positioning, resulting in less strain on the caregiver. 

The DPS is rated for 275 kg.

 

1.1

 

Purpose

 

The purpose of the DPS is to ease the transition of a suspended person from (a) sitting position 
to (b) lying down position. It also allows the caregiver to reposition the angle of the suspended 
patient while in the sling to facilitate a more comfortable and optimal user position. 

The DPS is suitable for all types of users, especially users having difficulty in reaching a 90-
degree sitting position, for example due to their medical condition.  

After thorough cleaning the DPS can be used by multiple users (See 4.1). 

NOTE:

 Do not use DPS for purposes other than originally intended. 

1.2

 

Safety instructions

 

 

Important - Warning! 
Read user manual and safety instructions before using the product.

 

 

NEVER use the DPS if there are any indications of wear or suspicion of other defects on 
the track systems, the Luna, accessories and the DPS itself. In such cases the DPS must be 
taken out of service immediately and you must contact your dealer or Winncare Nordic.  

 

Take the necessary steps to ensure that neither caregiver nor patient can be injured 
when carrying out lifts, transfers and tilts. Always check surroundings and move obstacles 
out of the way. 

 

Always ensure that the sling is attached correctly, before lifting a patient. 

 
NOTE:

 Read the user guide for Luna IPx4 before using the DPS.

 

1.3

 

Before use 

The caregiver or user is responsible for making the following checks before use: 

 

The caregiver has received instructions or read the user manual for the DPS. 

 

Ensure that the annual service check has been carried out. 

 

If in doubt about how to use the DPS please contact your dealer or Winncare Nordic for 
advice and guidance.

 

 

Summary of Contents for IPx4

Page 1: ...Please read this user manual carefully before using the product This user manual is intended as a guide for caregivers and other users as well as technical personnel during servicing and maintenance...

Page 2: ...2...

Page 3: ...___________ 7 3 2 Bed to chair__________________________________________ 7 3 3 Chair to bed _________________________________________ 7 4 MAINTENANCE______________________________________________ 8 4...

Page 4: ...ers See 4 1 NOTE Do not use DPS for purposes other than originally intended 1 2 Safety instructions Important Warning Read user manual and safety instructions before using the product NEVER use the DP...

Page 5: ...nt Contents of the box 1 The DPS 2 User manual 3 Label for hand control 1 6 Transport If the DPS is to be transported between sites it is recommended to use the original packaging In case of return sh...

Page 6: ...o the spreader bar attachment point 4 Connect the DC cable to the Luna The socket is located on the side opposite the hand control socket The DPS is ready for use if the instruction has been followed...

Page 7: ...hair When lifting a patient from a bed the DPS should be positioned in the lying position For a smooth transfer the caregiver can use the bed s adjustment functions if available Position the patient i...

Page 8: ...luding ceiling hoist It is the legal owner s duty to ensure that these service checks are carried out on the product The following items is to be inspected during the annual check Bolts o Check for lo...

Page 9: ...erly connected Attach DC cable as seen in 2 1 Check that the DC cable going into the actuator is connected Defective hand control Check if up down buttons on the hand control works If not try a differ...

Page 10: ...10 6 2 Dimensions Measurements All dimensions in mm Lying position Sitting position A Height 550 750 B Width 525 170 C Track to sling attachment point 864 1070 C A B...

Page 11: ...y damage the environment if not handled in an environmentally responsible way 7 2 Quality Winncare Nordic is certified according to the quality standard ISO 9001 and ISO 13485 These certifications mea...

Page 12: ...nufacturer Winncare Nordic ApS Taarnborgvej 12 C 4220 Korsoer Denmark Phone 45 70 27 37 20 Fax 45 70 27 37 19 E mail info Winncare dk www Winncare dk 93 42 E F CLASS I EN ISO 10535 EN 60601 1 2 910 00...

Reviews: