This assessment must take the following into account: 

the department’s surveillance capacities: A standardized protocol of the establishment for using the side 

rails may be drawn up. 
-  the  physical  and  mental  condition  of  the  patient:  his/her  needs,  capacities,  lucidity,  size,  agitation.  This 
assessment should be repeated at regular intervals. 
 

4.2. .Assembly of wooden barriers for junior bed A680-00 and adult beds A579-/A580-00 and 

A581-00 

 
WINNCARE wooden barriers

 

are designed to be used on all WINNCARE beds that have panels with recesses 

needed for their assembly 
 
 
 
 
The wooden barriers

 

are designed to be used on WINNCARE beds with a bed frame length of 1,700 mm for 

junior beds and 2,000 mm for adult beds that have panels with recesses needed for their assembly and that 
are not equipped with bed frame extensions. 
The barrier

 

has been tested in accordance with the test method of standards NF EN 50637 and 

NF EN 60601-2-52. These barriers

 

must be combined with mattresses whose technical characteristics are 

included in the bed’s instructions. 
 
For installation of the wooden fence, refer to the attached installation instructions. 
To facilitate the installation of the barrier, use the variable height of the bed up to a minimum height of 35 cm 
between the ground and the underside of the foot or head panel. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Mini 35

 cm

 

These barriers must be provided if specified in the medical prescription. The 
benefit/risk ratio must be reassessed every 24 hours. 
Barriers that go from the patient’s feet to their head are recommended for children 
who have lost their autonomy. 

 

INCOMPATIBLE SIDE BARRIERS CAN CREATE HAZARDS. 

Summary of Contents for A579-00

Page 1: ...Le Pas du Ch teau 85670 SAINT PAUL MONT PENIT TEL 33 0 2 51 98 55 64 FAX 33 0 2 51 98 59 07 Email info medicatlantic fr Site Internet https www winncare fr USER S MANUAL WOODEN BARRIERS A579 00 A580...

Page 2: ...Regulation EU 2017 745 relating to medical devices They are tested in conformity with standard EN 60601 2 52 2010 in their commercial configurations including the boards and accessories that we manufa...

Page 3: ...nt falls BUT it can pose a hazard injuries falling after trapping a limb in the side rails suffocation after trapping the head neck or chest A risk benefit assessment should preferably be carried out...

Page 4: ...that have panels with recesses needed for their assembly and that are not equipped with bed frame extensions The barrier has been tested in accordance with the test method of standards NF EN 50637 an...

Page 5: ...n in the low cross bar guide The assembly is identical for all full length wooden barriers and only the barrier of the A680 00 junior bed is composed of a low wood metal cross member to respect the st...

Page 6: ...6 The element composed of the wood and metal bar is always placed in the lower part The arrow down and to the inside of the bed 5...

Page 7: ...7 6 8 7 9 10...

Page 8: ...rrier To lower the side rail Take the upper bar and lift it up to its locking at the top of the panel Check if it is properly engaged To lower the side rail 1 l Lift the upper bar 2 Press the locking...

Page 9: ...ier gently down the bed make sure that no objects or parts of the patient s body or nursing staff are in the operating area until the stop is stopped The upper bar comes to rest on the lower metal woo...

Page 10: ...g used Side rails are a medical device and in this regard we inform you that it must not be modified under any circumstances You must ensure its traceability If you assemble different types of medical...

Page 11: ...itions of use and maintenance is 5 years for accessories especially for wooden side rails 5 6 Warranties All of our products carry a warranty against any manufacturing defect provided the normal condi...

Page 12: ...12...

Reviews: