background image

 

 
 
 
 
 
 
 
 
 

3. ASSEMBLY CONDITIONS 

 

3.1. General description 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

4. CONDITIONS FOR USE 

 

4.1. Function 

 
The side rails are designed to prevent patients from falling while asleep or being transported  
BUT are not designed to prevent a patient from leaving his or her bed voluntarily. Many accidents happen 
when the patient tries to leave his or her bed despite the side rails being in place. 
In some cases, the side rails helps to prevent falls, BUT it can pose a hazard: injuries, falling after trapping a 
limb in the side rails, suffocation after trapping the head, neck or chest. A risk/benefit assessment should 
preferably be carried out of the side rails to decide whether or not to use it. 
 
 
 

Observe the specified 

environmental conditions 

Cross beam

 

 

Sliding finger 

 

High cross-beam guide 
 

Low cross-beam guide 

Adult bed : section 92x29 Lg 2012 

Cross beam

 

Children bed : section 92x29 Lg 1712 

Cross beam

 

 

High cross-beam guide 
 

Cross beam

 

 

Low cross-beam guide 
 

Sliding finger 

 

Be sure to check the position of the panels and the correct sliding of the barrier. 

Summary of Contents for A579-00

Page 1: ...Le Pas du Ch teau 85670 SAINT PAUL MONT PENIT TEL 33 0 2 51 98 55 64 FAX 33 0 2 51 98 59 07 Email info medicatlantic fr Site Internet https www winncare fr USER S MANUAL WOODEN BARRIERS A579 00 A580...

Page 2: ...Regulation EU 2017 745 relating to medical devices They are tested in conformity with standard EN 60601 2 52 2010 in their commercial configurations including the boards and accessories that we manufa...

Page 3: ...nt falls BUT it can pose a hazard injuries falling after trapping a limb in the side rails suffocation after trapping the head neck or chest A risk benefit assessment should preferably be carried out...

Page 4: ...that have panels with recesses needed for their assembly and that are not equipped with bed frame extensions The barrier has been tested in accordance with the test method of standards NF EN 50637 an...

Page 5: ...n in the low cross bar guide The assembly is identical for all full length wooden barriers and only the barrier of the A680 00 junior bed is composed of a low wood metal cross member to respect the st...

Page 6: ...6 The element composed of the wood and metal bar is always placed in the lower part The arrow down and to the inside of the bed 5...

Page 7: ...7 6 8 7 9 10...

Page 8: ...rrier To lower the side rail Take the upper bar and lift it up to its locking at the top of the panel Check if it is properly engaged To lower the side rail 1 l Lift the upper bar 2 Press the locking...

Page 9: ...ier gently down the bed make sure that no objects or parts of the patient s body or nursing staff are in the operating area until the stop is stopped The upper bar comes to rest on the lower metal woo...

Page 10: ...g used Side rails are a medical device and in this regard we inform you that it must not be modified under any circumstances You must ensure its traceability If you assemble different types of medical...

Page 11: ...itions of use and maintenance is 5 years for accessories especially for wooden side rails 5 6 Warranties All of our products carry a warranty against any manufacturing defect provided the normal condi...

Page 12: ...12...