Document No.: TMM-2220-13
Revision: A
Page 24 of 30
SWITCH SCHEMATIC FOR “X” OPTION
Surgical support leg limit switch cord MUST be plugged in for device to function correctly. If surgical support leg
limit switch cord is unplugged or damaged the back & seat tilt motors will not actuate.
ALL EXCESS WIRE IS PLACED IN FRONT OF CONTROL BOX
GROUND WIRE ATTACHES TO GROUND POST
(“A” OPTION ONLY)
BATTERY CORD PLUGS INTO PORT #13
COLUMN PLUG (TMM5X) OR PLUGS (TMM5XT) PLUG INTO
PHONO TO DIN ADAPTER PLUGS THEN PLUG INTO PORTS
#1 (AND PORT #2 FOR TMM5XT)
LONGEST WIRE FROM WIRING ASSEMBLY IS THE BACK
SECTION ACTUATOR – PLUGS INTO PORT #3
MIDDLE LENGTH WIRE FROM WIRING ASSEMBLY IS THE
LEG SECTION ACTUATOR - PLUGS INTO PORT #4
SHORTEST LENGTH WIRE FROM THE WIRING ASSEMBLY
IS FOR SEAT TILT ACTUATOR – PLUGS INTO PORT #5
SUPPORT LEG LIMIT SWITCH CORD PLUGS INTO PORT #7
PENDANT CABLE PLUGS INTO PORT #9