Document No.: TMM-2220-13
Revision: A
Page 21 of 30
“F” OPTION: FOLDING FOOTREST
WHEN POSITIONING FOOTREST,
BE AWARE OF PINCH POINTS
To stow footrest, place both hands on
red handles and lift.
For patient comfort, stow footrest prior
to articulating chair into supine
(stretcher) position.
EXTENDED FOOTREST STOWED FOOTREST
“E” OPTION: EXPORT
“E” option designates chair was shipped with power
cord specific to receiving country.
“A” option power cord is directly attached to chair.
“B” option power cord is attached to battery charger
(# TMA57-15).
Power cords depicted in image may not match
power cord shipped with your chair.
OPERATING RED POWER SHUTOFF
BUTTON
•
When button is pressed, following items are
disabled:
o
Chair Articulation
o
Seat Height Adjustment
•
To reset red button, twist clockwise until button
pops back up.
EMERGENCY SHUTOFF PROCEDURE
During an emergency shutoff due to erratic performance, follow these steps:
1. Immediately press the RED emergency shutoff button on back of base cover to disable all electrical components.
2. Using proper safe patient handling procedures, assist patient with egress from chair.
3. Call TMM Service Department to report incident and to initiate investigation.
4. Discontinue chair use until cleared by TMM Service Department.