OWNERS OPERATING & MAINTENANCE MANUAL
TMM3 VIDEO FLUOROSCOPY STRETCHER-CHAIR
Page 14 of 28
Document #OM-TMM-2228-13
Revision: V
OPERATING INSTRUCTIONS: CONTINUED
•
STOWING SIDE RAIL
Once the rail is lowered (“down” position), push the rail to fully engage the side rail retainer clip.
•
RAISING SIDE RAIL
Lift side rail until it is in “up” position.
NOTE: Red rail release tab will engage (lock) automatically. Ensure side rail is secure by
pulling on rail after it is raised.
ADJUSTING SIDE RAIL ANGLE
1. Remove cushion and seat pan by removing four pan head screws.
NOTICE: Do not discard seat pan head screws, or
additional hardware removed during this step. Parts will
be reused after adjustment.
2. Loosen top jam nut.
3. Adjust as needed:
a. To move inward, rotate bottom jam nut as shown.
b. To move outward, rotate bottom jam nut as shown.
4. Once adjustment is complete, re-tighten top jam nut and
re-attach seat pan and cushion.
NOTICE
•
You do not need to hold bottom nut to
tighten top.
•
When adjusting, ensure bottom jam nut is contacting seat frame.
•
Ball joint head angle does not matter.
•
Prior to adjustment, pull outward on side rail in order for system to settle.
•
Ensure top jam nut engages at least two full threads.
•
Do not over tighten top jam nut (there is no need).
5. Reattach seat pan and cushion.