Winco S300 Stretchair Owners Operating & Maintenance Manual Download Page 6

IMPORTANT - PLEASE READ

5

1

Guidance and manufacturer’s declaration –electromagnetic emissions 

2

The Stretchair is intended for use in the electromagnetic environment specified below. The customer or the user of the 

Stretchair should assure that it is used in such an environment.  

3

Emissions test  

Compliance  

Electromagnetic environment – guidance  

4

RF emissions  

CISPR 11  

Group 1  

The Stretchair uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment.  

6

RF emissions  

CISPR 11  

Class B  

The Stretchair is suitable for use in all establishments, including 

domestic establishments and those directly connected to the 

public low-voltage power supply network that supplies 

buildings used for domestic purposes. 

7

Harmonic emissions 

IEC 61000-3-2  

Not Applicable 

8

Voltage fluctuations /

flicker emissions  

IEC 61000-3-3  

Not Applicable 

Recommended separation distances  

between portable and mobile RF communications equipment and  

the Stretchair

The Stretchair is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer

or the user of the Stretchair can help prevent electromagnetic interference by maintaining a minimum distance between portable and 

mobile RF communications equipment (transmitters) and the Stretchair as recommended below, according to the maximum output 

power of the communications equipment.  

Rated maximum output power of 

transmitter 

W

Separation distance according to frequency of transmitter 

m

150 kHz to 80 MHz  

80 MHz to 800 MHz  

800 MHz to 2,5 GHz  

P

d

2

.

1

=

P

d

2

.

1

=

P

d

3

.

2

=

0,01  

0.12 

0.12 

0.23 

0,1  

0.38 

0.38 

0.73 

1  

1.2 

1.2 

2.3 

10  

3.8 

3.8 

7.3 

100  

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

in meters (m) can be 

estimated using the equation applicable to the frequency of the transmitter, where 

P

is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter manufacturer.  
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.  
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from

structures, objects and people.  

Summary of Contents for S300 Stretchair

Page 1: ...PLEASE DO NOT DISCARD THIS MANUAL KEEP FOR FUTURE REFERENCE AND TRAINING S300 Stretchair S550 S750 S999 Bariatric Stretchair OWNER S OPERATING MAINTENANCE MANUAL...

Page 2: ...D FOR SHIPPING 3 CAREFULLY REMOVE ALL OF THE COMPONENTS FROM THE CARTON 4 SAVE ALL BOXES AND PACKING MATERIAL UNTIL AFTER YOU HAVE ASSEMBLED YOUR PRODUCT AND HAVE VERIFIED THAT ALL COMPONENTS ARE FUNC...

Page 3: ...rly if it has been dropped or damaged or dropped into water Contact your Winco authorized dealer 5 DO NOT carry or pull this product by power supply cord or use the cord as a handle 6 Keep the cord aw...

Page 4: ...r transporting in or with ANY type of vehicle or trailer Winco chairs have not been tested or approved for use by an occupant in any type of vehicle or trailer 8 NEVER use the chair arms or backrest a...

Page 5: ...ropean Union EU health safety environmental requirements Independent U S product safety testing certification by Nationally Recognized Testing Laboratory NRTL Products with the following markings are...

Page 6: ...air is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Stretchair can help prevent electromagnetic interference by m...

Page 7: ...ixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked wit...

Page 8: ...ferential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage...

Page 9: ...eight 53 5 22 Seat Height 22 28 Seat Width 31 72 Layflat Length 72 22 30 Vertical Lift Chair 22 30 30 38 Vertical Lift Layflat 30 38 303 lbs Chair Weight 312 lbs 750 lbs Weight Capacity 1000 lbs S300...

Page 10: ...een engaged return chair to it s lowest position and recline the Stretchair slightly using the handset control FIG 3 2 Unlock the wheels and secure the wheels in the steer lock position FIG 2A or 2B 4...

Page 11: ...using the handset control FIG 3 3 Completely lower the armrests FIG 1 4 Disconnect the patients safety belt 5 Carefully lift the patient s legs and return the footrests to their upright position FIG 1...

Page 12: ...ting Swing leg for Patient Comfort 1 The Stretchair MUST BE in the upright position 2 Depress the swing leg release handle and adjust to desired elevation FIG 5 Footrests can be pivoted to cushion up...

Page 13: ...e been locked Adjusting Armrests for Patient Comfort 1 Ensure that the patient s arms are not resting on the armrest prior to adjustment 2 Loosen the armrest knobs FIG 8 BOTH adjustment knobs must be...

Page 14: ...FIG 10 2 Align the battery guides FIG 11B with battery bracket and slide the battery connectors FIG 11C into the control box 3 Push opposite end of battery in and ensure that the battery release hand...

Page 15: ...to avoid oil spilliage 6 Carefully unscrew all 4 brass fittings from the Power Unit FIG 19 7 Dispose of the old Power Unit carefully 8 Cut black tubing to remove old ferrule and threaded fitting Cut...

Page 16: ...nting bolt and spacers FIG 22 6 Remove the lower mounting bolt and spacers FIG 22 7 Carefully remove the old Linak Actuator 8 Position the new Actuator and install the LOWER mounting bolt and spacers...

Page 17: ...NING INFORMATION VISIT http www wincomfg com permablok info and cleaning instructions MAINTENANCE GUIDE Swing leg Gas Spring Replacement 1 Fully elevate the legrests for easier access to the gas sprin...

Page 18: ...OUR PRODUCTS Other Quality Winco Furnishings You May Be Interested In Visit us online at www wincomfg com 18...

Page 19: ...19...

Page 20: ...ts attached to the frame such as upholstery foam casters mechanisms or related parts that are covered under above warranties Winco Mfg LLC shall not be liable for any labor shipping or any other costs...

Reviews: